Abstract
When abdomen and pelvic were scanned with 128 channel MDCT, the gonadal exposure dose was measured with and without gonadal shield and the obtained images were evaluated. As a result, during abdominal MDCT scan, the gonadal exposure dose was measured $16.5{\pm}0.5$ mGy when the gonad shield was not used, and it was $7.5{\pm}0.3$ mGy when the large gonad shield($650m^2$) was used, which showed the effect of reduction in the gonadal exposure dose by 54%. During pelvic MDCT scan, the gonadal exposure dose was $9.5{\pm}0.3$ mGy when the gonad shield was not used, and it was $2.8{\pm}0.2$ mGy when the large gonard shield($650m^2$) was used, which showed the effect of reduction in the gonadal exposure dose by 70%. The images were obtained when using the gonad shield and when not using it during MDCT scan, and as a result of analyzing them with 5-point Likert scale, in the abdominal image, it was 4.1 points irrespective of whether using the gonad shield or not. And also, in pelvic scan, it was 1.2 points when the gonad shield was used, and 4.1 points when it was not used. With the results above, it is considered that during the abdominal 128-MDCT scan, by using the gonad shield, the images should be obtained without being degraded and the exposure dose must be reduced.
최근 임상에서 사용하고 있는 128채널 MDCT의 복부 및 골반부 촬영 시 생식선 차폐 선량관리 자료로 활용하기 위하여 생식선의 피폭선량과 생식선 차폐 시 피폭선량을 측정하고 획득영상을 평가하였다. 평가한 결과 복부 MDCT 검사시생식선 차폐기구를 사용하지 않았을 경우에서 $16.5{\pm}0.5$ mGy를 나타냈으며, 대($650m^2$)크기의 차폐체를 사용하였을 경우 $7.5{\pm}0.3$ mGy로 측정되어 54%의 생식선 피폭선량 감소효과를 보였다. 골반부 MDCT 검사시 생식선 차폐기구를 사용하지 않았을 경우에서 $9.5{\pm}0.3$ mGy, 대($650m^2$)크기의 차폐체를 사용하였을 경우 $2.8{\pm}0.2$ mGy로 나타나 70% 의 생식선 피폭선량 감소효과를 보였다. MDCT를 이용한 검사 시 생식선 차폐기구 사용 유, 무에 따라 영상을 획득하여 Likert 5점 척도로 분석한 결과 복부 영상에서는 4.1점으로 동일한 결과를 보였다. 또한 골반부 검사에서는 생식선 차폐기구 사용하였을 때 1.2점, 생식선 차폐기구 사용하지 않았을 때 4.1점으로 나타났다. 이상의 결과에서 복부 128-MDCT 검사 시 생식선 차폐기구를 사용하여 영상의 질 저하 없이 영상을 획득하고 피폭선량을 감소하여야 할 것으로 사료된다.