DOI QR코드

DOI QR Code

Analysis of Image Distortion by Mandibular Arch Form in Cone Beam CT and Panoramic Image

Cone Beam형 CT와 파노라마 영상에서 하악궁의 영상 왜곡 분석

  • 정천수 (예닮치과병원) ;
  • 이기흔 (전북대학교 방사선과학기술학과)
  • Received : 2013.06.14
  • Accepted : 2013.08.25
  • Published : 2013.08.31

Abstract

This thesis intends to analyze tooth distortion by mandibular arch form by reproducing existing panorama image and reconstructed panorama image of Cone Beam CT data with the three-dimensional computer program. The diameter of tooth measured in Cone Beam CT's cross-section image and reconstructed panorama was synchronized without any big change from incisors to posteriors. But, panorama showed serious distortion as going to posteriors after showing a little distortion in incisors. The panorama reconstructed for patients' individual arch showed reduced distortion than panoramas used generally. In addition, panorama showed serious distortion from incisors to posteriors and it means that distortion is reduced in reconstructed panorama.

Cone Beam CT 데이터를 3차원 컴퓨터 프로그램을 이용하여 기존의 파노라마 영상과 재구성된 파노라마 영상을 재현해서 악궁형태에 따른 치아의 왜곡을 분석하고자 하였다. Cone Beam CT의 횡단면 영상과 재구성 파노라마에서 측정된 치아의 직경은 전치부에서 구치부로 갈수록 큰 변화없이 일치되는 것을 알 수 있었다. 하지만 파노라마는 전치부에서 약간의 왜곡이 보이다가 구치부로 갈수록 왜곡이 더 심한 것을 알 수 있었다. 환자 개개인의 악궁에 맞게 재구성된 파노라마가 일반적으로 사용되는 파노라마 촬영보다 왜곡이 더 감소한다는 것을 알 수 있었다. 또한 파노라마는 전치부에서 구치부로 갈수록 왜곡이 더 심해지는 것을 알 수 있었는데, 이 또한 재구성된 파노라마 영상에서는 왜곡이 감소한다는 것을 알 수 있었다.

Keywords

References

  1. C. H. Lim, S. C. Kim, H. Y. Jung, D. H. Hong, I. G. You, C. S. Jeong, "The Study for Radio Protection According to a Possible Danger of Exposure During dental X-ray Examination", J. Korean Soc. Radiol, pp. 238-239, October, 2011. https://doi.org/10.7742/jksr.2011.5.5.237
  2. C. S. Jeong, E. H. Mo, G. H. Lee, B. H. Han, S. C. Kim, J. S. Kim, C. H. Lim, "A Study on the Appropriate Reconstruction of the CBCT Images of Mandibular Canals", J. Korean Soc. Radiol, pp. 370-374, December, 2011. https://doi.org/10.7742/jksr.2011.5.6.369
  3. J. H. Lee, W. J. Han,, Y. H. Choi, E. K. Kim, "Measurement of maxillary sinus volume and available alveolar bone height using computed tomography", Korean J Oral Maxillofac. Radiol, pp.35-41, March, 2003.
  4. H. J. Jeon, "Diagnosis of Maxillary Sinus Septa using the CBCT and Accuracy of panoramic radiograph", published Master's dissertation, Ewha Womans University, pp 2-3, 2011.
  5. Y.S. Choi, H. E. Hwang, S. R. Lee, "Clinical application of cone beam computed tomography in dental implant", J Korean Dent Assoc, pp.172-179, March, 2006.
  6. H. M. Pinsky, S. Dydal, R. W. Pinsky, K. A. Misch, D. P. Sarment, "Accuracy of three-dimensional measurements using cone-beam CT", Dentomaxillofac Radiol, pp.410-416, November, 2006.
  7. Velasquez-Plata D, Lawrence R, Chris C, Marden E. "Maxillary sinus septa : a3-demensional computerized tomographic scan analysis", IntJOral MaxillofacImplants, pp.854-859, November, 2002.
  8. C. S, Cheong, B. H. Cho, D. S. Hwang, "Evaluationofmaxillary sinususing cone-beam CT in patients scheduled for dental implant in maxillary posteriorarea", J Korean Assoc OralMaxillofac Surg, pp.21-25, February. 2009
  9. J. H. Lee, W. J. Han,, Y. H. Choi, E. K. Kim, "Measurement of maxillary sinus volume and available alveolar bone height using computed tomography", Korean J Oral Maxillofac. Radiol, pp.35-41, March, 2003.
  10. S. I. Choi, "A study for the image quality enhancement methods of a dental digital panoramic imaging system", published Master's dissertation, Yonsei University, pp. 22-44, 2007.
  11. Farman AG, "ALARA still applies", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, pp.395-397, October, 2005.