Analysis and reduction of thermal magnetic noise in liquid-He dewar for sensitive low-field nuclear magnetic resonance measurements

  • Hwang, S.M. (Korea Research Institute of Standards and Science) ;
  • Yu, K.K. (Korea Research Institute of Standards and Science) ;
  • Lee, Y.H. (Korea Research Institute of Standards and Science) ;
  • Kang, C.S. (Korea Research Institute of Standards and Science) ;
  • Kim, K. (Korea Research Institute of Standards and Science) ;
  • Lee, S.J. (Korea Research Institute of Standards and Science)
  • Received : 2013.06.21
  • Accepted : 2013.06.26
  • Published : 2013.06.30

Abstract

For sensitive measurements of micro-Tesla nuclear magnetic resonance (${\mu}T$-NMR) signal, a low-noise superconducting quantum interference device (SQUID) system is needed. We have fabricated a liquid He dewar for an SQUID having a large diameter for the pickup coil. The initial test of the SQUID system showed much higher low-frequency magnetic noise caused by the thermal magnetic noise of the aluminum plates used for the vapor-cooled thermal shield material. The frequency dependence of the noise spectrum showed that the noise increases with the decrease of frequency. This behavior could be explained from a two-layer model; one generating the thermal noise and the other one shielding the thermal noise by eddy-current shielding. And the eddy-current shielding effect is strongly dependent on the frequency through the skin-depth. To minimize the loop size for the fluctuating thermal noise current, we changed the thermal shield material into insulated thin Cu mesh. The magnetic noise of the SQUID system became flat down to 0.1 Hz with a white noise of 0.3 $fT/{\surd}Hz$, including the other noise contributions such as SQUID electronics and magnetically shielded room, etc, which is acceptable for low-noise ${\mu}T$-NMR experiments.

Keywords

References

  1. J. Clarke, and A. Braginski, "The SQUID Handbook," Wiley-VCH, 2006.
  2. H. C. Seton , J. M. S. Hutchinson, and D. M. Bussel, "A 4.2 K receiver coil and SQUID amplifier used to improve the SNR of low-field magnetic resonance images of the human arm," Meas. Sci. Tech., vol. 8, pp. 198-207, 1997. https://doi.org/10.1088/0957-0233/8/2/015
  3. R. McDermott, N. Kelso, S-K. Lee, M. Mossle, M. Mück, W. Myers, B. ten Haken, H. C. Seton, A. H. Trabesinger, A. Pines and J. Clarke, "SQUID detected magnetic resonance imaging in microtesla magnetic fields," J. Low Temp. Phys., vol. 135, pp. 793, 2004. https://doi.org/10.1023/B:JOLT.0000029519.09286.c5
  4. G. Stroink and C. MacAulay, "Thermal magnetic noise generated by an eddy current shielded room," Rev. Sci. Instrum., vol. 57, no. 4, pp. 658-660, 1986. https://doi.org/10.1063/1.1138885
  5. S. Uzunbajakau, A. P. Rijpma, J. Dolfsma, H. J. G. Krooshoop, H. J. M. ter Brake, M. J. Peters, and H. Rogalla, "Magnetic flux fluctuations due to eddy currents and thermal noise in metallic disks," IEEE Trans. Magn., vol. 39, no. 4, pp. 2018-2023, 2003. https://doi.org/10.1109/TMAG.2003.812720
  6. N. Kasai, K. Sasaki, S. Kiryu, and Y. Suzuki, "Thermal magnetic noise of dewars for biomagnetic measurements," Cryogenics, vol. 33, pp. 175-179, 1993. https://doi.org/10.1016/0011-2275(93)90132-8

Cited by

  1. Study on elimination of screening-current-induced field in pancake-type non-insulated HTS coil vol.29, pp.3, 2016, https://doi.org/10.1088/0953-2048/29/3/035009