Simulation Study to Improve Ion Transmission Efficiency in Mass Spectrometry with a Dual Ion Funnel Ion Source

  • Baek, Sun Jong (Division of Mass Spectrometry Research, Korea Basic Science Institute) ;
  • Kim, Seung Yong (Division of Mass Spectrometry Research, Korea Basic Science Institute) ;
  • Kim, Taeman (MS Development Group, Genomic System Research Center, Bioneer Corporation) ;
  • Kim, Hyun Sik (Division of Mass Spectrometry Research, Korea Basic Science Institute)
  • Received : 2013.12.17
  • Accepted : 2013.12.24
  • Published : 2013.12.30

Abstract

We performed computer simulations to improve transmission efficiencies of a dual ion funnel system implemented on an FT-ICR MS. We found that the low m/z range from 50 to 150 could be significantly improved by operating the two ion funnels at different RF amplitudes. These new operational conditions could be applied to analyze metabolome samples, which require high sensitivity in the m/z range from 50 to 1,000.

Keywords

References

  1. Shaffer, S. A.; Tang, K.; Anderson, G. A.; Prior, D. C.; Udseth, H. R.; Smith, R. D. Rapid Communications in Mass Spectrometry 1997, 11, 1813 https://doi.org/10.1002/(SICI)1097-0231(19971030)11:16<1813::AID-RCM87>3.0.CO;2-D
  2. Kim, T.; Tolmachev, A. V.; Harkewicz, R.; Prior, D. C.; Anderson, G.; Udseth, H. R.; Smith, R. D. Anal. Chem. 2000, 72, 2247 https://doi.org/10.1021/ac991412x
  3. Kim, T.; Park, M. A. "Ion guide for mass spectrometers", US Patent 7,495,212.
  4. Cobb, J. S.; Easterling, M. L.; Agar, J. N. J. Am. Soc. Mass Spectrom. 2010, 21 (6), 949. https://doi.org/10.1016/j.jasms.2010.02.009
  5. Dahl, D. A. SIMION 3D (Version 8.1) Idaho National Engineering and Environmental Laboratory; Idaho Falls, ID, 2000.
  6. Tolmachev, A. V.; Kim, T.; Udseth, H. R.; Smith, R. D.; Bailey, T. H.; Futrell, J. H. Int. J. Mass Spectrom. 2000, 203, 31. https://doi.org/10.1016/S1387-3806(00)00265-7
  7. Kim, T. Ph.D. Thesis, 1997, p154, McGill University.