An Effective Block of Radioactive Gases for the Storage During the Synthesis of Radiopharmaceutical

방사성의약품 합성에서 발생하는 방사성기체의 효율적 차단

  • Chi, Yong Gi (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Kim, Dong Il (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Kim, Si Hwal (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Won, Moon Hee (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Choe, Seong-Uk (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Choi, Choon Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Seok, Jae Dong (Department of Nuclear Medicine, Samsung Medical Center)
  • Received : 2012.07.02
  • Accepted : 2012.09.21
  • Published : 2012.10.27

Abstract

Purpose : Methode an effective block was investigated to deal with volatile radioactive gas, short lived radioactive waste generated as a result of the routinely produced radiopharmaceuticals FDG (2-deoxy-2-[$^{18}F$]fluoro-D-glucose) and compound with $^{11}C$. Materials and Methods : All components of the radiation stack monitoring and data management system for continuous radioactive gas detection in the air extract system purchase from fixed noble gas monitor of Berthold company. TEDLAR gas sampling bags purchase from the Dongbanghitech company. TEDLAR gas sampling bags (volume: 10 L) connected via paraflex or PTFE tubing and Teflon 3 way stopcock. When installing TEDLAR gas sampling bags in Hot cell on the inside and not radioactive gas concentrations were compared. According to whether the Hot cell inside a activated carbon filter installed, compare the difference in concentration of the radioactive gas $^{18}F$. Comparison of radiation emission concentration difference of module a FASTlab and TRACElab. Results : Activated carbon filter are installed in the Hot cell, a measure of the concentration of radioactive gas was 8 $Bq/m^3$. Without activated carbone filter in the hot cell was 300 $Bq/m^3$. Tedlar bag prior to installation of the radioactive gases a measure of the concentration was 3,500 $Bq/m^3$, $^{11}C$ synthesis of the measured concentration was 27,000 $Bq/m^3$. After installed a Tedlar bag and a measure concentration of the radioactive gases was 300 $Bq/m^3$ and $^{11}C$ synthesis was 1,000$Bq/m^3$. Conclusion : $^{11}C$ radioactive gas that was ejected out of the Hot cell, with the use of a Tedlar gas sampling bag stored inside. A compound of 11C is not absorbed onto activated carbon filter. But can block the release out by storing in a Tedlar gas sampling bag. We was able to reduce the radiation exposure of the worker by efficient radiation protection.

방사성의약품 제조 시 휘발성 기체의 경우에 완전 차폐가 되지 않고, Hot cell 외부로 그리고 배기덕트를 통해 작업자에게 외부피폭은 물론 호흡을 통해 내부피폭을 가져오게 한다. 처음에는 Hot cell 자체의 배출구를 막아서 방사성기체를 차단하려하였으나 장치에 맞는 기체 밀폐형 댐퍼의 제작이 어렵고, 크기가 맞지 않아서 설치 후에 여전히 문제점이 개선되지 않았다. 그러나 Tedlar gas sampling bag의 사용으로 합성 장치의 가스 배출구를 연결하여 방사성 기체를 저장하고 10반감기가 지난 후에 배출함으로써 작업자의 피폭을 확연히 줄이게 되었으며 $^{18}F$ 방사성 기체는 Hot cell 배출구에 활성탄 필터를 연결하고 최종 배출구에 2차 활성탄 필터를 사용함으로써 배출되는 방사능 농도를 90% 이상 줄여주었다. 단 반감기의 핵종인 경우는 위와 같은 경우를 이용하여 다음날 작업을 할 수 있지만 반감기가 긴 핵종들 같은 경우는 다음날 처리 할 수 없는 문제점들이 발생한다. Decay tank의 추가적인 문제점들을 보완하거나 기체상의 여러 방사성 입자들을 포집 할 수 있는 물질들이 만들어져야 할 것이다. 현재 우리나라는 최종 배출 공기 중 방사능 농도만을 규제하고 있으나 유럽 같은 경우 일일 배출 양과 연간 배출도 규제를 하고 있다. 방사성의약품 합성 시 발생하는 많은 방사성 물질들을 보다 효과적으로 친환경적으로 처리할 수 있는 여러 연구들이 이루어져야 할 것이다.