The Suggestion of MF BIA Analyzer Method to Reflect the Resistance in the Extracelluar and Intracelluar Fluid at Acupoint & the Evaluation of Reflection Rate on the Tissue under Skin

MF-BIA Analyzer를 이용한 경혈의 세포 내액과 외액에 의한 저항성분 추정 방법 제시 및 반영률 평가

  • Kim, Soo-Byeong (The Eastern & Western Bio-Medical System Lab, Dept. of Biomedical Engineering, College of Health Science, Yonsei University) ;
  • Lee, Na-Ra (The Eastern & Western Bio-Medical System Lab, Dept. of Biomedical Engineering, College of Health Science, Yonsei University) ;
  • Kim, Won-Ky (Dept. of Biomedical Engineering, College of Health Science, Yonsei University) ;
  • Lee, Yong-Heum (The Eastern & Western Bio-Medical System Lab, Dept. of Biomedical Engineering, College of Health Science, Yonsei University)
  • 김수병 (연세대학교 보건과학대학 의공학과 동서의료시스템 연구실) ;
  • 이나라 (연세대학교 보건과학대학 의공학과 동서의료시스템 연구실) ;
  • 김원기 (연세대학교 보건과학대학 의공학과) ;
  • 이용흠 (연세대학교 보건과학대학 의공학과 동서의료시스템 연구실)
  • Received : 2012.12.04
  • Accepted : 2012.12.18
  • Published : 2012.12.27

Abstract

Objectives : The aim of this study was to suggest a new method to estimate the resistance by acupoint compositions by using the multi-frequency bioelectrical impedance analysis(MF-BIA) of 5 kHz, 50 kHz and 200 kHz within 2 cm of acupoints as a local segment. Moreover, we try to confirm the relation to between measured resistance at skin surface and measured resistance by various manual acupuncture needle insertion depth. Methods : Ten male subjects participated in this study. We measured the resistance at left/right ST36 at skin surface and various manual acupuncture needle insertion depth(skin, 5 mm, 10 mm, 15 mm, 20 mm, 25 mm and 30 mm). Results : It was also observed that the all measured resistances were the highest at 5 kHz and the lowest at 200 kHz. There were significant differences at 5 kHz, 50 kHz and 200 kHz between measured resistance at skin surface and measured resistance by various manual acupuncture needle insertion depth(p<0.05). There was no significant difference in the left and right identical acupoints under the identical condition(p>0.05). Conclusions : We conclude that the measured resistance at skin surface has limitation as to reflect the information of tissue. However, the measured resistance at each frequency was changed as similar pattern by different insertion depth. Hence, we confirmed the possibility of assumption on information of tissue which was expected to locate an acupoint.

Keywords

References

  1. Zhang D, Yan X, Zhang X, Liu C, Dang R, Xiao T, et al. Synchrotron radiation phasecontrast X-ray CT imaging of acupuncture points. Anal Bioanal Chem. 2011 ; 401(3) : 803-8. https://doi.org/10.1007/s00216-011-4913-7
  2. Yin L, Jin X, Qiao W, Sun J, Shi X, Tian J, et al. PET imaging of brain function while puncturing the acupoint ST36. Chin Med J. 2003 ; 116(12) : 1836-9.
  3. Voll R. Topographic positions of the measurement points in electroacupuncture. Am J Acupuncture. 1977 ; 5 : 97.
  4. Motoyama H. Electrophysiological and preliminary biochemical studies of skin properties in relation to the acupuncture meridian. Int Assoc for Religion and Parapsychology. 1980 ; 6 : 1-36.
  5. Zhu ZX. Research advances in the electrical specificity of meridians and acupuncture points. Am J Acupunct. 1981 ; 9 : 203-16.
  6. Kracmar F. Die biophysikalischen Grundlagen der Akupunktur. Dtsch Z Akup. 1962; 11(6): 131-40.
  7. Ionescu-Tirgoviste C, Constantin C, Bratu I. Electrical skin resistance in the diagnosis of neuroses. Am J Acupuncture. 1974 ; 2 : 247-52.
  8. Yan X, Zhang X, Liu C, Dang R, Huang Y, He W, et al. Do acupuncture points exist?. Phys Med Biol. 2009 ; 54(9) : 143-50. https://doi.org/10.1088/0031-9155/54/9/N01
  9. Myoung HS, Choi HS, Lee KJ, Lee YH. Development of acupoints discrimination system using electric properties of acupoints and non-acupoints. Conf Proc IEEE Eng Med Biol Soc. 2008 ; 2008 : 4924-7.
  10. Kim SB, Lee NR, Lee SW, Choi JY, Lee YH. Development of multi-frequency impedance measurement system for acupuncture points and preliminary report of measurement results. Korean Journal of Acupuncture. 2012 ; 29(1) : 71-81.
  11. Deurenberg P, Tagliabue A, Schouten FJ. Multi-frequency impedance for the prediction of extracellular water and total body water. Br J Nutr. 1995 ; 73(3) : 349-58. https://doi.org/10.1079/BJN19950038
  12. Lusseveld EM, Peters ET, Deurenberg P. Multifrequency bioelectrical impedance as a measure of differences in body water distribution. Ann Nutr Metab. 1993 ; 37(1) : 44-51. https://doi.org/10.1159/000177748
  13. McNeill G, Fowler PA, Maughan RJ, McGaw BA, Fuller MF, Gvozdanovic D, et al. Body fat in lean and overweight women estimated by six methods. Br J Nutr. 1991 ; 65(2) : 95-103. https://doi.org/10.1079/BJN19910072
  14. Cohen S, Popp FA. Biophoton emission of the human body. J Photochem Photobiol B. 1997 ; 40(2) : 187-9. https://doi.org/10.1016/S1011-1344(97)00050-X
  15. Jung HH, Woo WM, Yang JM, Choi C, Lee J, Yoon G, et al. Left-right asymmetry of biophoton emission from hemiparesis patients. Indian J Exp Biol. 2003 ; 41(5) : 452-6.
  16. Lukaski HC, Bolonchuk WW, Hall CB, Siders WA. Validation of tetrapolar bioelectrical impedance method to assess human body composition. J Appl Physiol. 1986 ; 60(4) : 1327-32.
  17. Elsen R, Siu ML, Pineda O, Solomons NW. Sources of variability in bioelectrical impedance determinations in adults. In: Ellis KJ, Yasumura S, Morgan WD, eds. In vivo body composition studies. London : Institute of Physical Sciences in Medicine. 1987 : 184-8.
  18. Gibney MA, Arce CH, Byron KJ, Hirsch LJ. Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: implications for needle length recommendations. Curr Med Res Opin. 2010 ; 26(6) : 1519-30. https://doi.org/10.1185/03007995.2010.481203
  19. Kim KY, Park MA. Study on somatotype and subcutaneous fat distribution of adult males in the twenties. Journal of the Ergonomics Society of Korea. 1997 ; 1 : 97-105.