Abstract
In this study, Geant4 based Monte Carlo simulations were carried out for medical linear accelerator. Modified Medical Linac2 toolkit was used for calculation. The energy spectrum, most probable energy and the photon mean energy compared with the published results using the EGS4 code. The results well agreed with published results. The calculated results of photon fluence, energy fluence and mean energy according to the radius from the centre of the beam were analyzed. Monte Carlo simulation using Medical Linac2 code is considered to be useful for analysis of medical linear accelerator. Because the calculated results varies depending on Physics List model for same head structure. It it important to choose the right model for research purpose. Monte Carlo simulation using GEANT4 Medical Linac2 is a valuable for any novice to adopt this code to the study related to 6 MV photon fluence from medical linear accelerator.
본 연구에서는 GEANT4 toolkit을 이용하여 의료용 선형가속기에 대한 몬테칼로(Monte Carlo) 전산모사를 하였다. Medical Linac2 예제를 수정해서 사용하였다. 에너지스펙트럼, 최빈에너지, 평균에너지를 EGS4 결과와 비교 하였고 선속의 중심에서부터 반경에 따른 단위면적당 광자수, 단위면적당 에너지, 평균에너지를 분석하였다. 그 결과 EGS4 결과와 큰 차이를 보이지 않기 때문에 Medical Linac2 예제의 선속특성에 관한 전산모사에 큰 문제점은 없는 것으로 판단된다. 같은 헤드구조에서도 Physics List의 모델에 따라서는 결과에 차이가 발생하므로 연구 환경에 알맞은 Physics List 모델을 선택하는 것이 중요하다고 판단된다. 본 연구는 처음 몬테칼로 전산모사를 접하는 사용자가 선속특성에 대한 전산모사를 수행하고 6 MV 광자선속의 특성을 분석하는 과정에 많은 도움이 될 것으로 사료된다.