Manufacture of Versatile Phantom for the Daily Quality Assurance of Radiation Therapy Equipments

방사선치료 장비들의 매일 정도관리를 위한 다용도 모형의 제작

  • Cha, Dong-Soo (Department of Radiation Oncology, Yeungnam University Hospital) ;
  • Lee, Jae-Seung (Department of Radiation Oncology, Good Samaritan Hospital)
  • 차동수 (영남대학교병원 방사선종양학과) ;
  • 이재승 (한동대학교 선린의료원 방사선종양학과)
  • Received : 2010.05.02
  • Accepted : 2010.05.28
  • Published : 2010.05.31

Abstract

We examined the accuracy and efficiency of phantom by applying the designed phantom in order to check daily quality assurance easily by objective criteria and to confirm daily quality assurance of linear accelerator, simulator, and CT-simulator. The results of 10 weeks of linear accelerator output dose using American Association of Physicists in Medicine(AAPM) daily quality assurance guide were measured within ${\pm}1%$ of error. Mechanical check of laser alignment, optical distance indicator(ODI), CT scanner laser and alignment of gantry lasers with the center of imaging plane were measured within ${\pm}1mm$. Daily average working time for daily quality assurance of radiation therapy equipments was 38 minutes. The designed phantom was easy to install and daily quality assurance was possible with only one installation. The aspects reproducibility and efficiency as well as accuracy of quality assurance were excellent.