Abstract
The aim of this study is to investigate PAE, as the result of the test of kVp accuracy, according to detector measurement method. Based on the indicated value of 70kVp, each distance between a focus and a kVp meter was 100cm, 80cm and 60cm and the angle of X-ray tube was set on $5^{\circ},\;10^{\circ},\;15^{\circ},\;20^{\circ},\;25^{\circ},\;30^{\circ}$. Each indicated value, 60kVp, 70kVp, 80kVp, 90kVp and 100 kVp, was used compare Small focus with Large focus. As a result, PAE on the side of cathode was higher than it on the side of anode in the case of 100cm and PAE on the side of anode was higher in the case of 80cm and 60cm. The coefficient rate was stable both the side of cathode and anode in the case of 100cm and it was fluctuated in the case of 80cm and 60cm. PAE in the case of Small focus was higher than Large focus and it was disproportionate to an indicated value. Error rate was in inverse proportion to the indicated value.
본 연구는 검출기 측정방법에 따라 관전압 정확도 검사결과인 PAE값이 어떻게 달라지는지 알아보고자 한다. 실험방법으로는 측정거리와 X선관 각도, 방향에 따른 PAE값의 변화실험에서는 초점과 측정기간의 거리를 100cm, 80cm, 60cm에 위치시킨 상태에서 관전압지시치는 70kVp, X선관 각도는 음극, 양극측으로 $5^{\circ},\;10^{\circ},\;15^{\circ},\;20^{\circ},\;25^{\circ},\;30^{\circ}$로 맞추어 측정하고, 관전압지시치에 따른 PAE값의 변화실험에서는 관전압 60kVp, 70kVp, 80kVp, 90kVP, 100kVp를 설정하고 소초점과 대초점을 전환하며 측정하였다. 측정거리와 X선관 각도, 방향에 대한 PAE값의 결과를 보면 100cm일 경우에는 음극측에서 전체적으로 PAE값이 높게 나타났으며 80cm, 60cm일 경우에는 양극측에서 높게 나타났다. PAE값의 변동율은 100cm일 경우 음, 양극측 모두 안정적으로 나타났으며 80cm, 60cm일 경우에는 불균형적으로 나타났다. 관전압지시치에 따른 PAE값의 결과로는 대초점보다는 소초점이 높게 나타났으며, 지시치가 클수록 높게 나타났다. 오차 범위는 지시치가 클수록 적게 나타났다.