Development of Electronic Circuit for Korotkoff Sounds Detecting Signal on Forearm Electronic Blood Pressure Monitor

팔뚝 전자혈압계의 코로트코프 음 신호 검출을 위한 전자 회로 개발

  • Received : 2009.12.15
  • Accepted : 2010.02.19
  • Published : 2010.03.30

Abstract

In this study, we developed a circuit device detecting korotkoff sounds of forearm electronic blood pressure monitor by digital signal. In order to test a circuit detecting signal from korotkoff sounds, systolic and diastolic pressure were compared our developed circuit device with the existing forearm electronic blood pressure monitor (Model: SE-7000, Korea). Devices for an experiment composed of a forearm cuff, a stethoscope, an amplifier, a PC with A/D board, etc. Results of korotkoff sounds was similar to a pattern of oscilometric signals from the existing forearm electronic blood pressure monitor. We thought it is possible to measure blood pressures, if blood pressures were detected precisely using signals of korotkoff sounds.

본 연구에서는 팔뚝 전자혈압계의 코로트코프 음을 디지털로 검출하는 신호검출회로 장치를 개발하였다. 본 연구에서 개발한 혈압계의 코로트코프 음 신호검출회로를 실험하기 위하여 기존의 팔뚝 전자혈압계 (Model: SE-7000, Korea)와 개발한 전자회로의 수축기 혈압과 이완기 혈압을 측정하여 비교 분석하였다. 실험을 위한 장비는 암형 커프, 청진기, 증폭기 및 A/D보드장착 PC 등으로 구성하였다. 본 연구 결과에서 코로트코프 음 신호는 기존의 팔뚝 전자혈압계에서 검출하는 오실로메트릭 신호 검출 패턴과 비슷한 경향을 보였다. 수은 혈압계와 차이에 의한 결과에서 코로트코프 음과 오실로메트릭 신호를 비교하면 수축기 혈압은 $6.75{\pm}2.02mmHg$로 나타났고, 이완기 혈압은 $7.24{\pm}3.40mmHg$ 로 나타났다. 코로트코프 음 신호를 이용한 혈압 측정치가 정밀하게 검출된다면, 기존의 팔뚝 전자혈압계를 정밀하게 검출 가능할 것으로 판단된다.