Reliability of measurement devices for measuring the ankle joint motion

발목 관절 가동범위 측정을 위한 측정도구의 신뢰도 연구

  • Hong, Wan-Sung (Department of Physical Therapy, Dong-Nam health college) ;
  • Kim, Gi-Won (Department of Physical Therapy, KeukDong college)
  • 홍완성 (동남보건대학 물리치료과) ;
  • 김기원 (극동정보대학 물리치료과)
  • Received : 2009.02.10
  • Accepted : 2009.05.10
  • Published : 2009.06.30

Abstract

Purpose: The purposes of this study were to establish the concurrent validity of the universal goniometer and electronic inclinometer for ankle joint of motion, and to determine the inter-tester and the intra-tester reliability of these two instruments. Methods: Subjects were instructed 25 healthy subjects. Ankle range of motion was measured on two separate occasions 2 or 3 days by two physical therapists. Ankle dorsiflexion and plantarflexion was by using an universal goniometer and an electronic inclinometer. Results: The Pearson product-moment correlation between the two instruments was 0.78~0.80. The ICCs for inter-tester reliability ranged from 0.63 to 0.73 for universal goniometer and ranged from 0.81 to 0.88 for electric inclinometer. The ICCs for intra-tester reliability showed a wide variation(ICC=0.61 to 0.86). Conclusion: These findings indicate that the two instruments are reliable instruments for measuring ankle joint range of motion. The results also indicate that the two instruments can be used interchangeably for measuring ankle motions.

Keywords