DOI QR코드

DOI QR Code

Micro-TEM Cell을 사용한 표준 전자기장의 발생 및 측정불확도 평가

Standard Field Generation Using a Micro-TEM Cell and Its Measurement Uncertainty Evaluation

  • 강진섭 (한국표준과학연구원 기반표준부 전자파센터) ;
  • 김정환 (한국표준과학연구원 기반표준부 전자파센터) ;
  • 강웅택 (한국표준과학연구원 기반표준부 전자파센터) ;
  • 강노원 (한국표준과학연구원 기반표준부 전자파센터) ;
  • 강태원 (한국표준과학연구원 기반표준부 전자파센터)
  • Kang, Jin-Seob (Center for Electromagentic Wave, Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS)) ;
  • Kim, Jeong-Hwan (Center for Electromagentic Wave, Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS)) ;
  • Kang, Ung-Taek (Center for Electromagentic Wave, Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS)) ;
  • Kang, No-Weon (Center for Electromagentic Wave, Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS)) ;
  • Kang, Tae-Weon (Center for Electromagentic Wave, Division of Physical Metrology, Korea Research Institute of Standards and Science(KRISS))
  • 발행 : 2009.01.31

초록

본 논문에서는 micro-TEM cell을 사용한 표준 전자기장 발생법을 기술하고 측정불확도를 평가하였다. 표준 전자기장 발생 시스템은 auto-leveling 기능을 가진 신호발생부, 최대 1.2 GHz까지 동작하는 micro-TEM cell, 서미스터 마운트를 사용한 전력측정부로 구성된다. 표준 전자기장 발생법의 타당성을 보이기 위해 $10\;MHz{\sim}1\;GHz$ 대역에서 전자기장의 세기 20 V/m에 대해 실시된 전자기장의 세기 국제비교(CCEM.RF-K20)의 참여 결과를 제시하였다.

In this paper, a standard field generation method using a micro-TEM ceil is described and its measurement uncertainty is evaluated. The standard field generation system consists of an auto-leveling signal source, a micro-TEM cell operating up to 1.2 GHz, and a power measuring Instrument using a thermistor mount. Measurement results of a field strength key comparison (CCEM.RF-K20) for the field strength of 20 V/m at frequencies between 10 MHz and 1 GHz are presented for validating the standard field generation method.

키워드

참고문헌

  1. IEC 61000-4-3:2006, Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test
  2. M. Kanda, R. D. Orr, "Generation of Standard Electromagnetic Fields in a TEM cell", Nat. Bur. Stand. Tech Note 1319, Aug. 1988
  3. K. Munter, R. Pape, and J. Glimm, "Portable E-field strength meter and its traceable calibration up to 1 GHz using a "μTEM" cell", IEEE Trans. Instrum. Meas., vol. 46, no. 2, pp. 549-550, Apr. 1997 https://doi.org/10.1109/19.571908
  4. No-Weon Kang, Jin-Seob Kang, Dae-Chan Kim, and Jeong-Hwan Kim, "Fabrication of small reference probe and its application", IEEE Trans. Instrum. Meas., vol. 56, no. 2, pp. 435-438, Apr. 2007 https://doi.org/10.1109/TIM.2007.890795
  5. No-Weon Kang, Jin-Seob Kang, Dae-Chan Kim, Jeong-Hwan Kim, and Joo-Gwang Lee, "Characterization method of electric field probe by using transfer standard in GTEM cell", to be published in IEEE Trans. Instrum. Meas., Apr. 2009 https://doi.org/10.1109/TIM.2008.2008592
  6. D. Pozar, Microwave Engineering, Addison-Wesley Publishing Company, 1990
  7. Fundamentals of RF and Microwave Power Measurements (Part 2), Agilent Application Note 1449-2, Jul. 2006
  8. 강진섭, 김정환, "Air line 임피던스 표준을 이용한 동축형 부하의 임피던스 정밀측정", 한국전자파학회논문지, 11(5), pp. 788-795, 2000년 8월
  9. 강진섭, 김정환, "두꺼운 외부 도체를 가진 손실이 있는 동축선로의 특성임피던스 근사", 한국전자파학회논문지, 11(4), pp. 649-656, 2000년 6월
  10. KRISS-99-070-SP, 측정불확도 표현지침, 한국표준과학연구원, 1999년 4월 수정판
  11. M. Kanda, D. Camell, Jan P. M. de Vreede, J. Achkar, M. Alexander, M. Borsero, H. Yajima, N. S. Chung, and H. Trzaska, "International comparison GT/RF 86-1 electric field strengths: 27 MHz to 10 GHz", IEEE Trans. on Electromagn. Compat., vol. 42, no. 2, pp. 190-205, May 2000 https://doi.org/10.1109/15.852413
  12. Final Report on GT-RF Key comparison CCEM. RF-K20: "Comparison of Electrical Field Strength Measurements", Aug. 2005