Abstract
The kerfs between arrayed piezoelectric elements in a medical ultrasonic transducer or a piezoelectric composite transducer are generally filled by polymeric materials. The boundary condition of the elements for lateral mode vibration is changed according to the kerf-filling materials, so that the resonance frequency for longitudinal mode of the transducer is also varied. In this study, to investigate the resonance frequency variation for an arrayed transducer experimentally, the piezoelectric vibration elements of $14mm{\times}0.22mm{\times}0.44mm$ were fabricated and those were linearly arrayed. And, the resonance frequencies were measured for three cases of kerf-filling condition, non-filling and two different kinds of epoxy filling. Conclusively, it is confirmed that the resonant frequency variation shows the similar tendency with the theoretical one for the longitudinal mode.
의료용 초음파 프로브나 압전복합재료 트랜스듀서 등에 있어서는 배열된 압전세라믹 요소간의 공간, 즉 kerf에 주로 폴리머계열의 물질을 충진하고 있다. 이 경우 압전 요소의 횡방향 진동모드의 경계조건이 바뀌어 실제 사용되는 종방향 진동모드의 공진 주파수에도 영향을 미치게 된다. 본 연구에서는 kerf충진 매질에 따른 공진주파수의 변화를 실험적으로 고찰하기 위하여 크기가 $14mm{\times}0.22mm{\times}0.44mm$인 PZT 압전요소의 선형배열에 의한 초음파 트랜스듀서를 제작하고, kerf충진 매질이 없는 경우와 특성이 다른 두 종류의 에폭시로 충진한 경우 각각에 대한 측정을 행하였다. 그 결과 kerf충진 매질에 따른 종방향 공진주파수의 천이 경향이 이론해석 결과와 유사함을 확인하였다.