Study on the Dose Characteristics of the PTW-LinaCheck Dosimeter and Its Application to Daily Output Measurement

PTW-LinaCheck 측정기의 선량 특성과 일일 출력측정 응용에 관한 연구

  • Jeong, Dong-Hyeok (Department of Radiation Oncology, School of Medicine, Wonkwang University Hospital) ;
  • Lee, Kang-Kyoo (Department of Radiation Oncology, School of Medicine, Wonkwang University Hospital) ;
  • Moon, Un-Chul (Department of Radiation Oncology, School of Medicine, Wonkwang University Hospital) ;
  • Kim, Hyun-Jin (Department of Radiation Oncology, School of Medicine, Wonkwang University Hospital) ;
  • Kim, Young-Seok (Department of Radiation Oncology, School of Medicine, Wonkwang University Hospital) ;
  • Moon, Sun-Rock (Department of Radiation Oncology, School of Medicine, Wonkwang University Hospital)
  • 정동혁 (원광대학교병원 방사선종양학과) ;
  • 이강규 (원광대학교병원 방사선종양학과) ;
  • 문언철 (원광대학교병원 방사선종양학과) ;
  • 김현진 (원광대학교병원 방사선종양학과) ;
  • 김영석 (원광대학교병원 방사선종양학과) ;
  • 문성록 (원광대학교병원 방사선종양학과)
  • Published : 2008.03.31

Abstract

In this study, we have investigated the dose characteristics of PTW-LinaCheck designed to detect output of medical LINAC and discussed clinical use of the detector. The reproducibility, linearity, and dose rate dependency of the dosimeter were measured for photons of 6 and 15MV and the electrons of 4, 6, 9, 12, and 16MeV. To know the error ranges of the measured data in daily output measurement, the response variations due to geometrical setup errors were measured. As a result of measurement, the error range from the geometrical setup and the reproducibility was less than ${\pm}0.6%$ for given beam qualities in daily output measurement, where the errors from the linearity and the dose rate dependency were negligible. Finally, we concluded that the LinaCheck dosimeter has a good characteristics in terms of dose and setup convenience in daily output measurement. In addition we have shown an examples of clinical use of this dosimeter for measuring daily output more than 60 days.

본 연구에서는 선형가속기 출력 검출용으로 상용화된 PTW-LinaCheck의 선량 특성을 조사하고 임상적 응용에 관하여 논의하고자 한다. 본 연구에서는 6, 15MV광자선 및 4, 6, 9, 12, 16 MeV전자선에 대하여 측정값의 재현성, 선형성, 선량률 의존성을 조사하였다. 또한 측정값의 오차범위를 조사하기 위하여 일일 점검 시 검출기의 셋업의 기하학적 오차에 의한 측정값의 변화를 조사하였다. 본 측정의 결과로서 일일 측정 시 셋업 오차에 의한 측정값의 변화는 재현성을 포함하여 ${\pm}0.6%$ 이내로 평가되었다. 이 때 선형성과 선량률 의존성에 의한 오차는 무시할 정도였다. 이를 통하여 LinaCheck는 선량 특성과 편리한 셋업의 관점에서 양호한 것으로 결론을 내렸다. 아울러 본 논문에서는 본 검출기를 사용한 60여 일간의 임상적 적용 사례도 제시하였다.