Abstract
When the calibration on Liquid Scintillation Counter using the Solid $^3H$ Standard Source of 200,000DPM is executed, the uncertainty due to activity and geometry difference, exists. Therefore, this paper intends to evaluate environmental samples comparatively accurately as decreasing this uncertainty existing in the process of calibration. For this, measurements on samples manufactured by $^3H$ Standard Source and sensitivity study were performed. Also, this paper verified calibration results using Radioactivity-Error-Analysis Method, and evaluated quantitatively the effect by geometry and activity difference based on verification result. According to the result of sensitivity study, in case of using the exposure time of 75 sec and Repeat method, the measuring accuracy and precision of about $1{\sim}3%$ were increased in comparison with the existing method. By analysis result, the effect by activity difference did not appear, and a plastic cell existing into Teflon vial made a role as reflector. The less the effect of plastic cells are decreased, the more activity is high, and the effect of those can be neglected at the activity of 200,000 DPM.
Activity가 20만 dpm인 고상 $^3H$ 표준선원을 사용하여 액체섬광계수기에 대한 교정을 수행할 경우 환경시료와는 Activity 및 Geometry 차이가 존재하고, 계측조건 차이로 인해 많은 불확실성이 존재할 수 있지만 이에 대한 연구결과가 없는 실정이다. 따라서 본 연구에서는 민감도분석을 통해 최적의 계측조건을 도출한 후 그 결과에 근거하여 Geometry 및 Activity 차이에 의한 영향을 정량적으로 평가하였고, 각 항목에 의한 영향이 나타날 경우 추가실험을 통해 원인을 규명하였다. 계측 결과에 대한 검증을 수행하기 위해 Chi-square test와 방사능오차분석을 수행하였고, 민감도분석 결과 본 연구에서 제안한 방법이 기존 방법에 비해 $1{\sim}3%$정도 오차가 감소하였다. 방사능오차분석 결과 Activity 차이에 의한 영향은 무시할 수 있었지만 Geometry 차이에 의한 영향이 크게 나타났고, 이에 대한 원인을 규명한 결과 비수용성인 플라스틱용기는 반사체 역할을 하였고, Activity가 높을수록 플라스틱에 의한 영향은 무시할 수 있었으며, 선원형태 차이에 의한 영향이 지배적인 것으로 나타났다.