DEVELOPMENT OF NANO-FLUID MOVEMENT MEASURING DEVICE AND ITS APPLICATION TO HYDRODYNAMIC ANALYSIS OF DENTINAL FLUID

미세 물 흐름 측정장치의 개발과 상아세관액의 수력학에의 응용

  • Lee, In-Bog (Department of Conservative Dentistry, School of Dentistry, Seoul National University) ;
  • Kim, Min-Ho (Department of Conservative Dentistry, School of Dentistry, Seoul National University) ;
  • Kim, Sun-Young (Department of Conservative Dentistry, School of Dentistry, Seoul National University) ;
  • Chang, Ju-Hea (Department of Conservative Dentistry, School of Dentistry, Seoul National University) ;
  • Cho, Byung-Hoon (Department of Conservative Dentistry, School of Dentistry, Seoul National University) ;
  • Son, Ho-Hyun (Department of Conservative Dentistry, School of Dentistry, Seoul National University) ;
  • Back, Seung-Ho (Department of Conservative Dentistry, School of Dentistry, Seoul National University)
  • 이인복 (치과보존학교실, 서울대학교 치의학대학원) ;
  • 김민호 (치과보존학교실, 서울대학교 치의학대학원) ;
  • 김선영 (치과보존학교실, 서울대학교 치의학대학원) ;
  • 장주혜 (치과보존학교실, 서울대학교 치의학대학원) ;
  • 조병훈 (치과보존학교실, 서울대학교 치의학대학원) ;
  • 손호현 (치과보존학교실, 서울대학교 치의학대학원) ;
  • 백승호 (치과보존학교실, 서울대학교 치의학대학원)
  • Published : 2008.03.31

Abstract

This study was aimed to develop an instrument for real-time measurement of fluid conductance and to investigate the hydrodynamics of dentinal fluid. The instrument consisted of three parts; (1) a glass capillary and a photo sensor for detection of fluid movement, (2) a servo-motor, a lead screw and a ball nut for tracking of fluid movement, (3) a rotary encoder and software for data processing. To observe the blocking effect of dentinal fluid movement, oxalate gel and self-etch adhesive agent were used. BisBlock (Bisco) and Clearfil SE Bond (Kuraray) were applied to the occlusal dentin surface of extracted human teeth. Using this new device, the fluid movement was measured and compared between before and after each agent was applied. The instrument was able to measure dentinal fluid movement with a high resolution (0.196 nL) and the flow occurred with a rate of 0.84 to 15.2 nL/s before treatment. After BisBlock or Clearfil SE Bond was used, the fluid movement was decreased by 39.8 to 89.6%.

본 연구의 목적은 나노리터 수준의 물 흐름을 계측할 수 있는 장치를 개발하고, 상아질의 물 투과성을 측정하여 치아과민증 치료제와 상아질 접착제의 상아세관 밀폐효과를 알아보고자 하였다. 본 연구에서 제작한 미세흐름 측정장치는 첫째, 물의 흐름을 감지하는 모세관과 광 센서부, 둘째 물의 흐름을 추적하는 서보모터와 구동부, 셋째, 모터의 회전을 측정하여 물의 이동량으로 변환하는 엔코더와 컴퓨터 기록장치 등 세 부분으로 구성되어 있다. 본 장비를 이용하여 교합 면이 절단되어 노출된 상아질의 물 투과도와 치아과민증 치료제인 BisBlock과 자가부식형 상아질접착제인 Clearfil SE bond의 상아질 밀폐효과를 측정하여 다음과 같은 결과를 얻었다. 1. 나노리터 수준의 물 흐름을 측정할 수 있는 장치를 제작하였고, 이를 이용하여 상아질의 물 투과도를 측정할 수 있었다. 2. 삭제 후 노출 연마된 상아질은 0.84 - 15.2 nL/s의 물 투과도를 보였고 Oxalate 제재인 BisBlock 이나 자가부식형 접착제 Cleafil SE bond 적용 시 투과도가 39.8 - 89.6% 감소하였다.

Keywords

References

  1. Roberson TM, Heymann HO, Swift EJ. art and science of operative dentistry. 4th edi. 14-62, Mosby, 2002
  2. Summitt JB, Robbins JW, Schwartz RS. Fundamentals of operative dentistry. 2nd edi. 8-22, Quintessence publishing co, Inc. 2001
  3. M Brannstrom. Dentin and pulp in restorative dentistry. Wolfe medical, 1981
  4. Hirata K, Nakashima M, Sekine I, Mukouyama Y, Kimura K. Dentinal fluid movement associated with loading of restorations. J Dent Res 70:975-978, 1991 https://doi.org/10.1177/00220345910700061301
  5. Orchardson R, Cadden SW. An update on the physiology of the dentin-pulp complex. Dent Update 28:200- 209, 2001 https://doi.org/10.12968/denu.2001.28.4.200
  6. Brannstorm M , Linden L, Johnson G. Movement of dentinal and pulpal fluid caused by clinical procedures. J Dent Res 5:679-682, 1968
  7. Pashley DH, Derkson GD, Tao L, Kalathoor S. The effect of a multi-step dentin bonding system on dentin permeability. Dent Mater 4:60-63, 1988 https://doi.org/10.1016/S0109-5641(88)80091-5
  8. Sauro S, Pashley DH, Montanari, Chersoni S, Carvalho RM, Toledano M, Osorio R, Tay FR, Prati C. Effect of simulated pulpal pressure on dentin permeability and adhesion of self-etch adhesives. Dent Mater 23:705-713, 2007 https://doi.org/10.1016/j.dental.2006.06.010
  9. Paphangkorakit J, Osborn JW. The effect of normal occlusal forces on fluid movement through human dentin in vitro. Arch Oral Biol 45:1033-1041, 2000 https://doi.org/10.1016/S0003-9969(00)00090-X
  10. De La Macorra JC, Escribano NI. Comparison of two methods to measure permeability of dentin. J Biomed Mater Res 63:531-534, 2002 https://doi.org/10.1002/jbm.10310
  11. Ozok AR, Wu MK, ten Cate JM, Wesselink PR. Effect of dentinal fluid composition on dentin demineralization in vitro. J Dent Res 83:849-853, 2004 https://doi.org/10.1177/154405910408301105
  12. Gregoire G, Joniot S, Guignes P, Millas A. Dentin permeability: self etching and one bottle dentin systems. J Prosthet Dent 90:42-9, 2003 https://doi.org/10.1016/S0022-3913(03)00258-0
  13. Pashley DH, Matthews WG, Zhang Y, Johnson M. Fluid shifts across human dentin in vitro in response to hydrodynamic stimuli. Arch Oral Biol 41:1065- 1072, 1996 https://doi.org/10.1016/S0003-9969(96)00059-3
  14. Ciucchi B, Bouillaguet S, Holz J, Pashley D. Dentinal fluid dynamics in human teeth, In vivo. J Endod 21:191-193, 1995 https://doi.org/10.1016/S0099-2399(06)80564-9
  15. Kolker JL, Vargas MA, armstrong SR, Dawson DV. Effect of desensitizing agents on dentin permeability and dentin tubule occlusion. J Adhes Dent 4:211-221, 2002
  16. Ratih DN, Palamara JEA, Messer HH. Dentinal fluid flow and cuspal displacement in response to resin composite restorative procedures. Dent Mater 23:1405-1411, 2007 https://doi.org/10.1016/j.dental.2006.11.029
  17. Ratih DN, Palamara JEA, Messer HH. Temperature change, dentinal fluid flow and cuspal displacement during resin composite restoration. J Oral Rehab 34:693-701, 2007 https://doi.org/10.1111/j.1365-2842.2007.01764.x
  18. Ciucchi B, Bouillaguet S, Delaloye M and Holz J. Volume of the internal gap formed under composite restorations in vitro. J Dent 25:305-312, 1997 https://doi.org/10.1016/S0300-5712(96)00032-2
  19. Martin RL Monticelli F, Brackett WW, Loushine RJ, Rockman RA, Ferrari M, Pashley DH, Tay FR. Sealing properties of mineral trioxide aggregate orthograde apical plugs and root fillings in an in vitro apexification model. J Endod 33:272-275, 2007 https://doi.org/10.1016/j.joen.2006.11.002
  20. Beneson W, Harris JW, Stocken H, Lutz H. Handbook of physics. P 153-210, springer, 2006
  21. Breschi L, Mazzoni A, Ruggeri A, Cadenaro M, Linarda RD, Dorigo EDS. Dental adhesion review: aging and stability of bonded interface. Dent Mater 24:90-101, 2008 https://doi.org/10.1016/j.dental.2007.02.009
  22. Tay FR, Pashley DH. Water treeing-a potential mechanism for degradation of dentin adhesives. Am J Dent 16:6-12, 2003

Cited by

  1. Real-time measurement of dentinal tubular fluid flow during and after amalgam and composite restorations vol.34, pp.6, 2009, https://doi.org/10.5395/JKACD.2009.34.6.467
  2. Real-time measurement of dentinal fluid flow during desensitizing agent application vol.35, pp.5, 2010, https://doi.org/10.5395/JKACD.2010.35.5.313
  3. Effect of oral health-related factors on oral health knowledge, attitude, and practice of college students vol.42, pp.4, 2018, https://doi.org/10.11149/jkaoh.2018.42.4.124