Progress in Medical Physics (한국의학물리학회지:의학물리)
- Volume 19 Issue 4
- /
- Pages.305-312
- /
- 2008
- /
- 2508-4445(pISSN)
- /
- 2508-4453(eISSN)
Analysis of Accuracy of Apparent Activity According to Calibration Method for High Dose Rate Brachytherapy Source
측정용 전리함에 따른 고 선량율 근접치료용 방사성 선원의 겉보기 활성도의 정확도 비교 분석
- Huh, Hyun-Do (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Choi, Jin-Ho (Department of Radiation Oncology, Gachon University of Science and Medicine) ;
- Choi, Sang-Hyoun (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Kim, Seong-Hoon (Department of Radiation Oncology, College of Medicine, Hanyang University) ;
- Kim, Woo-Chul (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Kim, Hun-Jeong (Department of Radiation Oncology, College of Medicine, Inha University) ;
- Lee, Re-Na (Department of Radiation Oncology, School of Medicine, Ewha Womans University) ;
- Kim, Kum-Bae (Department of Radiation Oncology, Korea Institute of Radiological and Medical Sciences) ;
- Hong, Seong-Eon (Department of Radiation Oncology, School of Medicine, KyungHee University) ;
- Shin, Dong-Oh (Department of Radiation Oncology, School of Medicine, KyungHee University)
- 허현도 (인하대학교 의과대학 방사선종양학교실) ;
- 최진호 (가천의과학대학교 방사선종양학교실) ;
- 최상현 (인하대학교 의과대학 방사선종양학교실) ;
- 김성훈 (한양대학교 의과대학 방사선종양학교실) ;
- 김우철 (인하대학교 의과대학 방사선종양학교실) ;
- 김헌정 (인하대학교 의과대학 방사선종양학교실) ;
- 이레나 (이화여자대학교 의과전문대학원 방사선종양학교실) ;
- 김금배 (원자력의학원 방사선종양학과) ;
- 홍성언 (경희대학교 의학전문대학원 방사선종양학교실) ;
- 신동오 (경희대학교 의학전문대학원 방사선종양학교실)
- Published : 2008.12.31
Abstract
The aim of this study is to introduce the accuracy of Ir-192 source's apparent activity using the well-type chamber and the Farmer-type ionization chamber in the high dose rate brachytherapy. We measured the apparent activity of Ir-192 that each medical center in the country has and the apparent activity of calibration certificate provided by manufacturer is compared with that by our experimental measurement. The number of sources used for the activity comparison was 5. The accuracy of the measured activity was in the range of -2.8% to -1.0% and -2.1% to 0.2% for the Farmer-type chamber system (Jig) and for the well-type, respectively. The maximum difference was within 1.0% for comparison with two calibration's tool. Our results demonstrate that well-type chamber as wall as Farmer-type chamber is a appropriate system as the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, by periods, a source calibration should be carried out.
본 연구에서는 측정 방법이 다른 두 전리함을 이용하여 고선량율 근접치료용 방사성 선원의 겉보기 활성도를 측정하여 선원의 활성도 정확성 및 두 장비간의 정확도를 비교 분석하였다. 국내 5개 의료기관을 선정하여 공기 중 기준거리에서 겉보기 활성도를 측정하고 동일한 선원에 대하여 측정 방법이 다른 두 팬톰을 동시에 측정하여 측정방법에 따른 상호 오차와 측정용 전리함의 정확도를 평가하였다. 우물형 전리함과 Jig 팬톰의 상대 오차는 각각 -2.1%~0.2%, -2.8%~-1.0%로 나타났다. 동일한 방사성선원에 대하여 두 교정 장비간의 정확도 평가는 최대 상대오차 1%로 나타났다. 선원의 세기 검증 시 측정 방법에 따른 의존도는 없었다. 고 선량율 근접방사선치료에 사용되는 방사성 선원의 세기의 변화는 환자의 종양 및 인접장기에 미치는 영향이 매우 크므로 방사성 선원의 세기 오차는 환자치료의 최적화에 심각한 영향을 미칠 수 있다. 그러므로 새로운 선원 교체 시 뿐만 아니라 주기적인 선원 점검이 요구된다.
Keywords