A Study on Calibration Procedures for Ir-192 High Dose Rate Brachytherapy Sources (고선량률(HDR) 근접치료의 동위원소 Ir-192에 대한 측정방법에 관한 고찰)

  • Baek, Tae-Seong;Lee, Seung-Wook;Na, Soo-Kyong
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.19 no.1
    • /
    • pp.19-26
    • /
    • 2007
  • Purpose: To compare of the accuracy among various measurement procedures of HDR Brachytherapy, and to evaluate the clinical suitability and usefulness of alternative PMMA (polymethylmethacrylateplastics: $C_5H_8O_2$) plate phantom without any additional cost due to the purchase of measuring apparatus. Materials and Methods: We made a comparative study on three types of measuring systems: well type chamber, source calibration jig, and PMMA plate phantom. Farmer type chamber was used for source calibration jig method and PMMA plate phantom method. Measurement was done 5 times each in comparison with the measurement values from manufacturer. Measurement results from experiment were compared with that from the manufacturer which is offered with the source whenever a source is substituted by a new one and evaluate the accuracy of source activity. Results: As a consequence of Ir-192 source measurement using well type chamber, source calibration jig and PMMA plate phantom, RMS (Root Mean Square) values for the relative error are 0.6%, 1.57%, 2.1%, respectively, compared with the data from manufacturer. And the mean errors with standard deviation are given $-0.2{\pm}0.5%$, $0.97{\pm}1.23%$, $-0.89{\pm}1.87%$ respectively. Conclusion: From the results shown by the three types of measurement system (well type chamber, source calibration jig, and PMMA plate phantom), the measurement with well type chamber produced the best accuracy. It turns out that we can also use the alternative system of PMMA plate phantom clinically without purchasing any additional particular apparatus since the system does not exceed the recommendation of AAPM (American Association of Physicists in Medicine), which requires the error range of within ${\pm}5%$.

  • PDF

Comparison of Calibration Methods of $^{192}\textrm{Ir}$ Sources for High Dose Rate Brachytherapy (고선량률 근접조사치료용 이리듐-192 방사성동위원소의 교정방법 비교연구)

  • Huh Hyun Do;Park Sung Yong;Lee Rena J;Shin Dong Oh;Kwon Soo Il;Loh John J K;Choi Jinho
    • Progress in Medical Physics
    • /
    • v.15 no.4
    • /
    • pp.192-196
    • /
    • 2004
  • The activity of Ir-192 sources for high dose rate (HDR) Brachytherapy in Korea were measured by using the well-type chamber and using the calibration Jig with the Farmer-type ionization chamber to compare the manufacturer certificated source strength which is supplied with each new Ir-192 source. The activity of two different source models used in six hospitals were measured. The range of measured activities to the manufacturer's suggested ones was -2.40% to +3.31% for the calibration Jig and -3.12% to 0.00% for the well-type chamber system. The source strength values given by the manufacturer for the 6 sources were within ${\pm}5%$ for the two different measuring equipment. Our results demonstrate that well-type chamber as wall as Farmer-type chamber system are appropriate system for the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, a source calibration should be carried out.

  • PDF

Analysis of Accuracy of Apparent Activity According to Calibration Method for High Dose Rate Brachytherapy Source (측정용 전리함에 따른 고 선량율 근접치료용 방사성 선원의 겉보기 활성도의 정확도 비교 분석)

  • Huh, Hyun-Do;Choi, Jin-Ho;Choi, Sang-Hyoun;Kim, Seong-Hoon;Kim, Woo-Chul;Kim, Hun-Jeong;Lee, Re-Na;Kim, Kum-Bae;Hong, Seong-Eon;Shin, Dong-Oh
    • Progress in Medical Physics
    • /
    • v.19 no.4
    • /
    • pp.305-312
    • /
    • 2008
  • The aim of this study is to introduce the accuracy of Ir-192 source's apparent activity using the well-type chamber and the Farmer-type ionization chamber in the high dose rate brachytherapy. We measured the apparent activity of Ir-192 that each medical center in the country has and the apparent activity of calibration certificate provided by manufacturer is compared with that by our experimental measurement. The number of sources used for the activity comparison was 5. The accuracy of the measured activity was in the range of -2.8% to -1.0% and -2.1% to 0.2% for the Farmer-type chamber system (Jig) and for the well-type, respectively. The maximum difference was within 1.0% for comparison with two calibration's tool. Our results demonstrate that well-type chamber as wall as Farmer-type chamber is a appropriate system as the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, by periods, a source calibration should be carried out.