CYCLIC FATIGUE OF THE SODIUM HYPOCHLORITE TREATED AND /OR STEAM AUTOCLAVED NICKEL-TITANIUM ENDODONTIC FILES

차아염소산나트륨 용액과 고압증기멸균이 근관치료용 니켈-타이타늄 파일의 주기적 피로 파절에 미치는 영향

  • Cho, Hye-Young (Department of Conservative Dentistry, College of Dentistry, Yonsei University) ;
  • Jung, Ii-Young (Department of Conservative Dentistry, College of Dentistry, Yonsei University) ;
  • Lee, Chan-Young (Department of Conservative Dentistry, College of Dentistry, Yonsei University) ;
  • Kim, Eui-Seong (Department of Conservative Dentistry, College of Dentistry, Yonsei University)
  • 조혜영 (연세대학교 치과대학 보존학교실) ;
  • 정일영 (연세대학교 치과대학 보존학교실) ;
  • 이찬영 (연세대학교 치과대학 보존학교실) ;
  • 김의성 (연세대학교 치과대학 보존학교실)
  • Published : 2008.01.31

Abstract

The purpose of this study was to determine the effect of sodium hypochlorite and steam autoclaving on the cyclic fatigue of nickel-titanium endodontic files. Two types of files with a .06 taper and #30 were used, $K3^{(R)}$ (SybronEndo, Glendora, California, USA) and Hero $642^{(R)}$(Micro-Mega, BesanCon, France). The files were divided into 6 experimental groups containing 10 files each group depending the soaking time in 6% sodium hypochlorite solution and number of cycles of steam autoclave. After sterilization, a cyclic fatigue test was performed on each file, and the fracture time was recorded in seconds. The control group underwent the cyclic fatigue test only. After the test, the surface characteristics of the files were observed using scanning electron microscopy (SEM). All groups containing the Hero 64~ files showed a similar cyclic fatigue fracture time. However, the cyclic fatigue fracture time with the $K3^{(R)}$ files was significantly shorter in groups which were treated with sodium hypochlorite than in the control group (P < 0.05), SEM revealed both Hero $642^{(R)}$ and $K3^{(R)}$ files to have significant corrosion on the file surface in groups treated with sodium hypochlorite, compared with the sharp and regular blades of the control group. $K3^{(R)}$ files showed more corrosion than the Hero $642^{(R)}$ files. Bluntness of the blades of the $K3^{(R)}$ file was observed in groups treated with steam autoclave. Although there was no obvious destruction on the surface of steam autoclaved Hero $642^{(R)}$ files, slight bluntness was observed. Sterilizing with a steam autoclave is much less destructive to $K3^{(R)}$ files than sodium hypochlorite. The longer time exposed to sodium hypochlorite, the more destructive pattern was shown on the blades of the files. Therefore, when using sodium hypochlorite solution, the exposure time should be as short as possible in order to prevent corrosion and increase the cyclic fatigue fracture time.

본 연구에서는 차아염소산나트륨 용액 및 고압증기멸균이 근관 치료용 Ni-Ti 파일의 주기적 피로 파절에 미치는 영향을 알아보고자 하였다. 파일 첨단 크기가 #30이고 .06 taper를 가지는 Hero $642^{(R)}$$K3^{(R)}$파일 70개를 10개씩 7개의 군으로 나누었으며, 6% 차아염소산나트륨 용액에 15분 (N15군), 30분 (N30군), 고압증기멸균은 $121\;^{\circ}C$의 온도, 0.12 MPa의 압력에서 20분간 시행하는 것을 일 회로 하여 5회 (H5군), 10회 (H10), 고압증기멸균 5회 후 차아염소산나트륨 용액 15분(H5N15), 고압증기멸균 10회 후 차아염소산나트륨 용액 30분 (H10N30) 처리 후 주기적 피로 파절 실험을 수행하여 파일이 파절될 때까지의 시간을 초 단위로 측정하였다. 대조군은 전처리를 하지 않고 주기적 피로 파절을 시행하였다. 95%의 신뢰도로 one-way analysis of varience를 이용하여 통계 처리하였으며, 다중 비교는 Tukey방법으로 분석하였다. 파일의 표면을 주사전자현미경으로 관찰하였다. 주기적 피로 파절에 소요된 시간은 Hero $642^{(R)}$는 각 군별로 유의할 만한 차이가 없었으며, $K3^{(R)}$는 N15, N30, H5N15군에서 대조군에 비해 짧은 시간이 소요된 것으로 나타났다 (p < 0.05). 주사전자 현미경에서 관찰했을 때 Hero $642^{(R)}$$K3^{(R)}$모두 N15, N30군에서 파일의 표면과 날이 부식이 되어 대조군에서 관찰된 일정 방향으로 배열된 정형화된 예리한 빗살무의가 부분적으로 파괴된 형태가 관찰되었으며, N15군에 비해 N30군에서 파일의 형태가 파괴된 양상이 뚜렷했으며, Hero $642^{(R)}$보다 $K3^{(R)}$가 심하게 부식된 양상이었다. H5군과 H10군에서는 $K3^{(R)}$ 파일은 날 부분이 무디어진 양상을 보였고, Hero $642^{(R)}$는 형태 파괴가 뚜렷하지 않았으나 H10군에서 표면의 빗살무늬가 약간 무디어진 양상을 보였다. H5N15군과 H10N30군에서는 그다지 뚜렷한 형태변화는 없었다. 본 연구 결과 $K3^{(R)}$파일을 사용 시에는 가능한 한 차아염소산나트륨에 적은 시간 노출시키는 것이 파일의 부식으로 인한 형태의 손상을 방지할 수 있고, 주기적 피로 파절에 걸리는 시간도 감소시킬 수 있을 것이다.

Keywords

References

  1. Walia H, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod 14:346-351, 1988 https://doi.org/10.1016/S0099-2399(88)80196-1
  2. Glosson CR, Haller RH, Dove SH, Del Rio CE. A comparison of root canal preparations using Ni-Ti hand, Ni-Ti engine-driven, and K-Flex endodontic instruments. J Endod 21(3):146-151, 1995 https://doi.org/10.1016/S0099-2399(06)80441-3
  3. Gambill JM, Alder M, Del Rio CE. Comparison of nickel- titanium and stainless steel hand-file instrumentation using computed tomography. J Endod 22(7):369- 375, 1996 https://doi.org/10.1016/S0099-2399(96)80221-4
  4. Thompson SA, Dummer PMH. Shaping ability of ProFile .04 taper series 29 rotary nickel-titanium instruments in simulated root canals. Part 1. Int Endod J 30:1-7, 1997 https://doi.org/10.1111/j.1365-2591.1997.tb01093.x
  5. Serene TP, Adams, JD, Saxena A. Nickel-Titanium instruments : Applications in Endodontics, St Louis : Ishiyaku EuroAmerica, Inc., 1995
  6. Pruett JP, Clement DJ, Carnes DLJr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 23(2):77-85, 1997 https://doi.org/10.1016/S0099-2399(97)80250-6
  7. Dieter GE. Mechanical metallurgy, 3rd ed., McGraw- Hill, New York, NY, 1986
  8. Silvaggio J, Hicks L. Effect of heat sterilization on the torsional properties of rotary nickel-titanium endodontic files. J Endod 23(12):731-734, 1997 https://doi.org/10.1016/S0099-2399(97)80344-5
  9. Stokes O, Fiore P, Barss J, Koerber A, Gilbert J, Lautenschlager E. Corrosion in stainless steel and nickel-titanium files. J Endod 25(11):731-735, 1999 https://doi.org/10.1016/S0099-2399(99)80119-8
  10. Halkel Y, Serfaty R, Wilson P, Speisser J, Alleman C. Mechanical properties of nickel-titanium endodontic instruments and the effect of sodium hypochlorite treatment. J Endod 24(11):731-735, 1998 https://doi.org/10.1016/S0099-2399(98)80163-5
  11. Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol 32(2):271-275, 1971 https://doi.org/10.1016/0030-4220(71)90230-1
  12. American Dental Association. Council on dental materials and devices. specification No. 28 for endodontic files and reamers. J Am Dent Assoc 93(8):831-817, 1976
  13. Halkel Y, Serfaty R, Bateman G, Senger B, Alleman C. Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instrument. J Endod 25(6):434-440, 1999 https://doi.org/10.1016/S0099-2399(99)80274-X
  14. Yared G, Dagher F, Machtou P. Cyclic fatigue of Profile rotary instruments after simulated clinical use. Int Endod J 32:115-119, 1999 https://doi.org/10.1046/j.1365-2591.1999.00201.x
  15. Yared G, Dagher F, Machtou P. Cyclic fatigue of Profile rotary instruments after clinical use. Int Endod J 33:204-207, 2000 https://doi.org/10.1046/j.1365-2591.1999.00296.x
  16. Fife D, Gambarini G, Britto L. Cyclic fatigue testing of Protaper NiTi instruments after clinical use. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 97:251-256, 2004 https://doi.org/10.1016/j.tripleo.2003.08.010
  17. Turpin YL, Chagneau F, Vulcain JM. Impact of two theoretical cross-sections on torsional and bending stressess of nickel-titanium root canal instrument models. J Endod 26(7):414-417, 2000 https://doi.org/10.1097/00004770-200007000-00009
  18. Gambarini G. Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and hightorque endodontic motors. J Endod 27(12):772-774, 2001a https://doi.org/10.1097/00004770-200112000-00015
  19. Li U, Lee B, Shih C, Lan W, Lin C. Cyclic fatigue of endodontic nickel-titanium rotary instruments : static and dynamic tests. J Endod 28(6):448-451, 2002 https://doi.org/10.1097/00004770-200206000-00007
  20. Gambarini G. Cyclic fatigue of Profile rotary instruments after prolonged clinical use. Int Endod J 34:386-398, 2001b https://doi.org/10.1046/j.1365-2591.2001.00259.x
  21. Thierry B, Tabrizian M, Savadogo O, Yahia L'H. Effect of sterilization process in NiTi alloy : surface characterization. J Biomed Mater Res 49(1):88-98, 2000 https://doi.org/10.1002/(SICI)1097-4636(200001)49:1<88::AID-JBM11>3.0.CO;2-I