The Noise Evaluation for Ragius 150 CR System

Regius 150 Computed Radiography 시스템의 Noise 평가에 관한 연구

  • Published : 2006.12.30

Abstract

The Noise of CR Systems is made up of X-ray quantum mottle and additional Imaging plate's structure noise, photon noise of lumine cence, noise of electrometer, quantization noise etc. In this Regius 150 system, SNR was increased from 8.2 to 30 with linearily according to radiation dose from 0.1 mR to 100 mR. It means that the Regius 150 system has enough trustability because of SNR is over 5 by Rose Model. NPS was calculated two dementional Fourier Transform with shake of pixel value in the white Image. In the spatial frequence range of $0.5\;lp/mm{\sim}2.5\;lp/mm$, the NPS was distributed $10^{-4}{\sim}10^{-5}$ at 1 mR X-ray dose. That is similar result compare other systems to that of Kodak CR system reported by Carlu, HR-CR system reported by Dobbins.

CR 시스템의 잡음(Noise)의 주요한 원인은 X-선 양자 모틀이며 그 외에 Imaging plate의 구조 노이즈, 휘진발광체의 광양자 노이즈, 전기계의 노이즈, 양자화 노이즈 등이 있다. Regius 150시스템에서 선량이 $0.1\;mR{\sim}100\;mR$까지 변화할 때 SNR 변화는 0.1 mR에서 8.2부터 선형성을 가지고 증가하고 있다. 이때 최저 SNR 값이 5 이상 되므로 충분한 신뢰성을 확보했다고 할 수 있다. 의료영상시스템의 가장 일반적인 노이즈 평가방법인 Noise Power Spectrum(NPS)을 이산적 화상데이터로부터 얻어진 Pixel 값의 요동으로부터 직접 푸리에 변환하여 구하였다. 주파수 $0.5\;lp/mm{\sim}2.5\;lp/mm$ 구간의 NPS은 선량이 많아질수록 개선되며 1 mR 조사 시 $10^{-4}{\sim}10^{-5}$ 사이에 분포하여, 시판되고 있는 타 회사의 CR시스템과 유사한 노이즈 특성을 가지고 있다.