Abstract
The present study was intended to examine a basic scheme to determine the biochemical oxygen demand(BOD) of livestock wastewater by means of six individual dissolved oxygen(DO) sensors and its multi-measurable meter. Maximal point of the first order time derivative of the DO difference between DO distribution of sterilized livestock wastewater and that of non-sterilized livestock wastewater, was considered as the oxygen uptake rate(OUR) of microorganisms in livestock wastewater, as determined to be 0.00074 mg $O_2/{\ell}{\cdot}sec$. The present study showed that there was a fair linear relationship(97.72%) between maximal OUR and BOD values of livestock wastewater, the latter being determined by classical Winkler azide method. It was thus concluded that the present multi-biosensor system might be applicable to an on-line system for measurement of BOD of livestock wastewater.
용존산소 센서 및 시스템에서 손실되는 산소량을 보정하기 위해 멸균처리 기법을 도입하였다. 멸균처리 시료의 DO 거동과 비멸균처리 시료의 DO 거동차(${\Delta}$DO)는 시스템의 영향과 무관하게 폐수 자체의 순수한 미생물만이 소모하는 산소변화량을 구하였다. 용존산소 센서로 축산폐수의 시간에 따른 DO거동(-d${\Delta}$DO/dt)을 측정하여 구한 폐수 자체의 미생물에 의한 산소소모속도는 0.00074mg $O_2/{\ell}{\cdot}sec$이었다. 축산폐수의 원액을 희석하여 다양한 BOD값을 갖도록 시료를 제조한 후 DO meter로 측정하여 구한 DO 변화량은 동일한 시료를 5일 BOD 측정방법인 Winkler Azide화 변법으로 구한 BOD값과 선형구간($30{\sim}60$분)에서 97.72%의 높은 상관 선형성을 보였다. 따라서 본 연구의 multi-biosensor 시스템은 축산폐수의 BOD를 짧은 시간에 정확하게 측정할 수 있는 가능성을 제시하였다.