Development of Standard Experimental Apparatus for PIV Performance Evaluation

PIV 성능시험을 위한 표준실험장치 개발

  • Published : 2006.12.30

Abstract

An experimental apparatus for PIV performance evaluation has been developed. Stardard uncertainty of a two-dimensional cross-correlation PIV system was investigated based upon the standard experimental apparatus, which was devised to model the rigid body rotating flows. For the systematic analysis of the uncertainty introduced by each component (algorithm, CCD camera, frame grabber) of the PIV system, standard images are fed into the component independently. The standard experiments show that 53% of the uncertainty in the present PIV system results from the frame grabber but the errors from the algorithm and digital camera are ignorable.

Keywords