Real-time Pesticide Assay on Live Tissue Using Electrochemical Graphite Pencil Electrode

살아있는 세포에서 전기화학적 흑연 연필심 전극을 사용한 살균제의 실시간 분석

  • Published : 2006.06.20

Abstract

A simply prepared graphite, pencil-type working electrode was utilized to monitor fenitrothion concentrations, using the cyclic voltammetry (CV) and square-wave (SW) stripping voltammetry methods. The optimum conditions for analysis were sought. A very low detection limit was obtained compared to that obtained when other common voltammetry methods are used. The optimal parameters of the pencil-type electrode were found to be as follows: a pH of 3.7, a frequency of 500 Hz, an SW amplitude of 0.1 V, an increment potential of 0.005 V, an initial potential of -0.9V, and a deposition time of 500 sec. The analytical detection limit was determined to be 6.0 ngL-1 (2.16410-11 molL-1) fenitrothion at SW anodic and CV, and the relative standard deviation at the fenitrothion concentration of SW anodic 10 ugL-1 was 0.30% (n = 15) under the optimum conditions. Analysis was directly conducted through in-vivo real-time assay.

간편하게 만들어진 흑연연필심 작업전극을 사용한 순환전압전류법과 사각파형 벗김 전압 전류법으로 살충제 페니트로치온을 정량하였다. 최적분석조건을 연구한 결과 기존의 전기화학적 분석법들 보다 정밀하고 낮은 검출한계에 도달하였으며 이들 조건은 다음과 같다. 수소이온농도: 3.7 pH, 벗김 사각파형 주파수: 500 Hz, 벗김 사각파형 증폭률: 0.1V, 벗김파형 상승전위: 0.005V, 석출전위: -0.9V, 석출시간: 500초에서 벗김전압전류법과 순환전압전류법의 페니트로치온 농도 검출한계는: 6.0 ngL1 (2.164×1011 molL-1) 이었다. 상대표준편차는 10 ugL-1을 15번 반복측정하여 0.30%의 정밀도 였으며. 위 최적 조건에서 실시간 살아있는 세포에서 분석 응용하였다.

Keywords

INTRODUCTION

Fenitrothion (MW=277.24) and its analogous pesticide are important in medicinal chemistry, and for humans. Fenitrothion can behave as an androgen agonist; can cause major changes in the hematological indices in warm-blooded animals, as well as cancer, Parkinson’s Disease, cell tumors, and other diseases1-5 and can be a cause of worry for many consumers of plant products such as olive oil because of its acute toxicity.6-9 Thus, various convenient methods for analyzing fenitrothion and its analogs have been developed, such as online reversed-phase liquid chromatography-gas chromatography (RPLC-GC),8 solid-phase extraction,10 a liquid chromatographic multi-residue method,11 gas chromatography ion-trap mass spectrometry detection (GC-MS),12 and gas chromatography (GC).9 These methods are very sensitive, but they require a compressing system as well as temperature-controlling systems, separation systems, and other spectrophotometric or electric detection systems. Recently, there has been an increasing demand for more sensitive and simpler methods of analysis.

Square-wave voltammetry techniques are very useful and popular for trace pesticides13-15 since they are compact, efficient, and sensitive.16,17 Various voltammetry solutions have been found to have low detection limits, which are required for pesticide analysis, depending on the working electrode systems used. For example, an acetycholine (ACh) biosensor fabricated with gienzymes/poly (diallyldimethyl ammonium chloride) (PDDA) multiplayer, film-modified platinum (Pt) electrodes through a layer-by-layer technique (LBL) attained a detection limit of 0.001 ugmL-1 for trichlorfon;18 a cobalt phthalocyanine-modified carbon ink electrode attained the detection limit of 28.8 nM for nabam;19 and an enzyme biosensor based on acetylcholinesterase (AChE) and choline oxidase (ChO) immobilized on the supported monomolecular layer attained the detection limit of 1.3×10-3 ppb for the dimethyl-2. 2-dichlorovinylphosphate (DDVP), 0.01 ppb for carbonfuran, and 0.03 ppb for escrine.20 Furthermore, the detection limit of isolated photosystem II (PSII) immobilized between the dialysis membrane and the Teflon membrane of the Clark oxygen electrode was 5×10-10 M for diuron;21 and a screen-printed enzyme electrode attained the detection limit of 5 nM (1.3 ppb) for trichlorfon.22 All these methods attained very low detection limits.

There are many articles about various kinds of pesticides, but hardly any article on the detection of fenitrothion using electrode systems could be found. As such, the detection and analysis of fenitrothion were conducted in this study, specifically using the pencil-type electrode, which can be easily prepared, and is renewable and low-cost.23,24 The pencil-type electrode is versatile for electrochemical measurements,25 and is able to detect the low concentration and/or volume of an analyte because it has a larger active electrode surface area.26 Various researches have been conducted using the pencil-type electrode or the pencil-type graphite electrode (PGEs).27,28 The use of this method enabled detection limits for fenitrothion lower than those mentioned above to be obtained in vivo. It was first applied to a direct assay in living tissue.

 

EXPERIMENT

Apparatus Reagents and the Procedure

Voltammetric systems were conducted using the CHI-660A Instruments Electrochemical Workstation (CH Instruments Inc., Cordova TN, USA). Three-electrode cell systems were used to monitor the cyclic and square-wave stripping voltammograms. The pencil-type electrode was prepared as a working electrode (Pentel Hi-Polymer Ain, 0.9 mm× 60 mm, Pentel Co., Ltd., Japan) using a 0.5-mm-diameter copper wire connected to the measurement system. Graphite nanotube paste electrode29-31 and DNA-mixed nanotube paste electrode32 were prepared. Ag/AgCl electrode (Cl coated 0.2 mm diameter wire type) and a platinum wire (0.2 mm diameter) served as the reference and auxiliary electrodes, respectively. All the solutions were prepared from double-distilled water (-18-M Ohm cm-1). 0.1-M NH4H2PO4, with a pH level of 3.7, served as a supporting electrolyte solution. All the other reagents were of analytical grade. The three-electrode system was immersed in a stirred solution of a known amount of fenitrothion. Prior to stripping, pre-concentration was carried out using dissolved oxygen. The common parameter for CV was a scan rate of 3.5 Vs-1, and the common parameters for stripping voltammetry were used under optimized conditions with both working electrodes. The voltammetric response of fenitrothion is dependent on the electrolyte solutions being used and on the pH. Various types of electrolyte solutions were tested, and from among these, the phosphate solution was found to yield the best results. In the electrode comparison at Fig. 1(a), the pencil-type electrode responded more sensitively than the bare-type nanotube paste electrode and the DNA-composed nanotube paste electrode in a CV peak height for oxidation and reduction scan. The results that were obtained by the baretype nanotube are as follows: 4.378×10-7 A (ox) and 8.821×10-7 A (Re); and by the DNA nanotube paste electrode: 4.137×10-6 A (Ox) and 1.144×10-7 A (Re), using 20-mgL-1 fenitrothion, under the same electrolyte conditions. Considerable noise was produced when the bare-type nanotube electrode was used in CV, and the anodic peak had a kind of peak current, while the cathodic peak was broad at the DNA nanotube paste electrode in CV.

Fig. 1(a) Cyclic voltammograms of graphite pencil electrode in 0.1M ammonium phosphate acid solutions (pH=3.69) containing of the 5-, 10-, 15-, 30-, 45-, 60-, 75-, 90-, and 105-mgL-1 fenitrothion, sacn rate of 3.5 V/s, and electrode comparison (insertion curve). (b) Square-wave stripping peak current of the 5-, 10-, 15-, 30-, 45-, 60-, 75-, 90-, and 105-mgL-1 fenitrothion concentrations, conditions of a pH 3.7, a frequency of 500 Hz, an SW amplitude of 0.1 V, an increment potential of 0.005 V, an initial potential of -0.9 V, and a deposition time of 100 sec.

 

RESULTS AND DISCUSSION

Fenitrothion concentration using cyclic and square-wave stripping voltammetry

First, the cyclic voltammetric peak potentials with a high concentration of 5- to 105-mgL-1 fenitrothion were studied. Fig. 1a shows the raw voltammograms of the results in 0.1-M NH4H2PO4, with a pH of 3.7, and 3.5 V/s scan rate, based on the cyclic voltammetric signals tested on the penciltype electrode. Between the initial cathodic scan of blank and 105-mgL-1 fenitrothion concentrations, a reduction peak appeared at 0.0 V. Then, at the next reverse scan, an oxidation peak of 0.3 V appeared, which was also well extracted from the background discharge and from the oxidation peak. That was the result of the spiked fenitrothion concentration between 5 and 105 mgL-1 at reduction and oxidation. At these conditions, the linear working equations of reduction y=0.7819 x -1.7533; a correlation of R2 = 0.9646 (y: peak current in A; x: concentration in mgL-1); and oxidation y = 1.8465 x -3.2016 and R2 = 0.9668 were obtained. These peak potentials were used for the determination of squarewave stripping voltammetry. Fig. 1b illustrates the SW signals from the fenitrothion concentration between 5 and 105 mgL-1, as in Fig. 1a. In the SW graph, the peak currents were gathered closely, sharply, and clearly at around 0.3 V. Based on the bigger graph with a line in Fig. 1b, it can be said that the response of the pencil-type electrode to fenitrothion is effective. The regressive equation was y=0.0182 x +2.0493, with a correlation of R2 = 0.993. The power of the peak currents gradually increased from 2.083 to 3.866×10-5 A, with the same conditions as those in the above detection. Through two detections of fenitrothion using CV and SW, respectively, the penciltype electrode was able to detect the pesticide in the mgL-1 concentration.

Experimental optimization for various parameters of square-wave stripping voltammetry

In Fig. 2a, various pH effects were examined using a 0.1-M ammonium phosphoric acid solution and 50-mgL-1 fenitrothion spiking. Based on a graph with a curve in the figure, the maximum peak currents (8.557×10-6 A) were obtained at a pH of 6.26. But the pH of 6.26 could not be an optimum pH because the peak potential was not 0.3 V at the SW graph in Fig. 2a. Only two peak currents (pH of 3.71 and 4.78) appeared at 0.3 V, but the pH of 3.71 had a more powerful peak current (5.198×10-6 A) than the pH of 4.78 (4.526×10-6 A) in the penciltype electrode. Thus, pH 3.7 was chosen as the optimum pH. At this condition, other parameters are used with Fig. 1(b) conditions. All the other experiments were performed using this pH of the electrolyte solutions. In Fig. 2b, various square-wave peak currents for the increment potentials of 0.005~0.040 V are shown within fixed 50-mgL-1 fenitrothion concentrations. Under these conditions, the accumulation time was fixed at 2 sec. The maximum peak current was obtained at 0.005 V. In the graph with a curve in the figure, the third point, 0.015 V, had the strongest peak current (8.621×10-6 A). The positions of the peak currents for the increment potential, however, were moving according to the SW graph in Fig. 2b; as such, the first increment potential, 0.005 V, was determined to be the optimum increment potential (8.377×10-6 A) at the pH of 3.7, which was detected in the prior experiment. The succeeding examinations conducted in the search for optimum conditions were carried out using these detected optimum parameters. After that, the optimum accumulation potential for fenitrothion was examined. The results of this examination are shown in Fig. 2c. Within the range of -0.9 to -1.6 V, the most powerful peak current (9.882×10-6 A) appeared at -1.3 V, according to the one-curve graph. At the first point, however, it is -0.9 V, and it had the sharpest SW signal, which was detected only at the potential of 0.3 V, at 8.377×10-6 A. Thus, -0.9 V was selected as the optimum initial potential. This potential was also used in the other examinations. Fig. 2d shows the effects of the SW accumulation time within a range of 10-500 sec under the same conditions and at a fixed fenitrothion concentration of 50 mgL-1. As expected, the peak current increased as the time increased. The heights of the peak currents for the pencil-type electrode made up an almost-straight line. The SW signals were also straight and sharp. From among the time parameters, the last point of the one-curve graph (500 sec) had the tallest peak current (9.327×10-4 A). However, at the range of 200~500 sec, two interference peak ions appeared with 0.0V and -0.4 V, which manifested only in high accumulation times, and the curve of that time parameter was the deepest and was observed at 0.3 V. It was decided that the optimum time parameter would be 500 sec. At every measurement step, the electrode surface was cleaned using weighing papers, and electrode stability was obtained with statistical precisions of within 5% error in each step, with each electrode usable in one or two months.

Fig. 2(a) Different pH effects of 2.68, 3.19, 3.71, 4.78, 5.68, 5.99, 6.26, and 6.56 on a 50-mgL-1 fenitrothion concentration, with electrolytes of the 0.1-M NH4H2PO4 solution. (b) Square-wave stripping voltammetric peak current at the various increment potentials of 0.005, 0.010, 0.015, 0.020, 0.025, 0.030, 0.035, and 0.040 V. (c) Various SW accumulation potentials of -0.9, -1.0, -1.1, -1.2, -1.3, -1.4, -1.5, and -1.6 V. (d) Various SW accumulation time of 10, 20, 50, 100, 200, 300, 400, and 500 sec in a 50-mgL-1 fenitrothion.

Concentration, statistical results, test on interference, and application

As shown in Fig. 3a, various detectable working concentration ranges were examined. After the background current was subtracted from the measured currents, those currents were plotted. A linear range within the low concentration range of 10-180 ngL-1 was exhibited. Within this range, anodic directions sharply appeared, with a regression equation of y=0.0347x+1.577 (y: current in A; x: concentration in ngL-1), with R2=0.9938 (R2: correlation coefficient). This equation can be used to determine the fenitrothion concentrations in pharmaceutical samples. The curve in the bigger graph in the figure is similar to a straight line. This means that the response of the pencil-type electrode is effective enough to detect an infinitesimal fenitrothion concentration, such as “ngL-1.” The precision of 15 times of the 10-ugL-1 solution yielded a relative standard deviation of 0.29%. Although the results are not shown here, they are highly reproducible. The peak currents varied between 1.824 and 5.224×10-4 A under the following optimum conditions: an initial potential of -0.9 V, a final potential of 1.1 V, an increment of 0.005 V, an amplitude of 0.1 V, a frequency of 500 Hz, a deposition time of 100 sec, and a sensitivity of 1e-4 AV-1. Since they gradually increased, they generated an almost-straight line. Thus, it was found that using the standard addition method in determining fenitrothion concentration could correct the interference of other species. At this condition, analytical detection limit was obtained for 6.0 ngL-1, with a 0.29% relative error, based on the signal-to-noise (S/N) characteristics of the data for the S/N = 3 given optimum conditions.

Various possible interference species were tested under optimum conditions with other pesticides and heavy metals, in a fixed fenitrothion concentration of 1 mgL-1 and 2 mgL-1, respectively. The proportion between fenitrothion and other substances was 1:10. The pesticides that were tested were chlorothalonil, dichlorvos, simazine, thiram, and diazinon. The heavy metals that were tested were Ba, Cr, Ce, Hg, Ge, Fe, and Cd. In these tests, 0.1-M ammonium phosphoric acid solution was used as the electrolyte solution. In the preceding order, they interfered with fenitrothion 9.36%, 5.48%, 3.56%, 0.19%, and 2.27%, with a pH of 3.70; and 7.41%, -11.16%, 22.27%, 58.77%, 3.51%, 19.19%, and 58.27%, with a pH of 3.70, respectively. These percentages indicate the increase in the fenitrothion peak current. The most interfering pesticide was chlorothalonil (2.372×10-4 A), and the most interfering heavy metal was Cd (6.399×10-4 A). These interfering substances can be corrected using the standard addition method. Using an electrode that is effective in detecting fenitrothion, two applications were conducted. The first was the examination of the pesticide in a live plant (Monochoria korsakowii Regel et Maack). After spiking 20 mgL-1 of fenitrothion in the water where the plant had been raised, the water and the plant were left alone for about a day. A higher concentration, 50-mgL-1 fenitrothion, was then used to spike the same water, and about another day was allowed to pass. About two days later, fenitrothion was directly detected in the live plant with peak current of 0.24 (±0.0026)×10-6 A (RSD 6.05%) at a repeatability rate of 5. The working electrode was inserted in the plant without destroying or damaging its tissue using three wire-type electrode systems. The results of this procedure are shown in Fig. 3b. As in all the detection procedures that had been conducted, the peak current was observed at around 0.3 V, with an initial potential of -0.9 V, a final potential of 1.1, an increment of 0.01 V, an amplitude of 0.015A, a frequency of 556 Hz, and a sensitivity of 1e-4 AV-1. This indicates that the detection of fenitrothion is possible with the use of the pencil-type electrode. The next application was carried out using a live carp brain with an initial potential of -0.9 V, a final potential of 1.1 V, an increment of 0.01 V, an amplitude of 0.15, a frequency of 556 Hz, a deposition time of 150 sec, and a sensitivity of 1e-6 AV-1. Preparation for this detection procedure was the same as that in the application using a live plant. A total of 70-mgL-1 fenitrothion was used to spike the water where a carp had been living. Three electrodes (standard electrode, auxiliary electrode, and working electrode) were implanted in both sides of the carp’s chest and head, respectively. The results of this procedure are also shown in Fig. 3c, specifically peak potential was obtained with the same peak sharp at repeatability rate of 3. The peak current shown in the figure is around 0.3 V, which is the same as in the other detection procedures conducted about two days later. The SW signal was obtained directly from the live fish’s brain. However, fenitrothion peaks in any other pond could not be obtained.

Fig. 3(a) SW stripping voltammograms of fenitrothion at various concentrations of 10, 20, 40, 60, 80, 100, 125, 154, and 180 ngL-1. (b) An application using a live plant, with different fenitrothion concentrations of 20 and 50 mgL-1. (c) An application using live fish brain, with different concentrations of 20 and 50 mgL-1.

 

CONCLUSIONS

The optimum conditions for determining and analyzing fenitrothion concentrations were sought using the pencil-type electrode, whose signal was shown to be more sensitive than that when common voltammetry methods are used. Optimum conditions derived from the experimental results were as follows: a pH strength of 3.70 with 0.1-M NH4H2PO4; an SW frequency of 500 Hz; an SW amplitude of 0.1 V; an increment potential of 0.005 V; an initial accumulation potential of -0.9 V; and a deposition time of 500 sec. A low range of 10~150 ngL-1 fenitrothion was obtained. Under the optimum conditions for analysis, the detection limit was determined to be 6.0 ngL-1, with a 0.29% relative error. It can be said that the method used in this paper is much more effective than other methods used in monitoring fenitrothion because detection was carried out with no direct destruction of tissue. The results of this study can find application in clinical science, electrochemistry, and other fields.

References

  1. Baldwin, R. T.; Martin, S. P. Toxico. Applied Pharmaco. 2004, 199, 118 https://doi.org/10.1016/j.taap.2003.12.029
  2. Leandro, C. C.; Hancock, P.; Fussell, R. J.; Keely, B. J. J. Chromatogra. A. 2006, 1103, 94 https://doi.org/10.1016/j.chroma.2005.10.077
  3. Caboni, P.; Sherer, T. B.; Zhang, N.; Taylor, G.; Na H. M.; Greenamyre, J. T.; Casida, J. E. Chem. Res. Toxicol. 2004, 17, 1540 https://doi.org/10.1021/tx049867r
  4. Chen, Z.; Robison, L,; Giller, R.; Krailo, M.; Davis, M.; Davies, S.; Shu, X. O. Int. J. Hyg. Environ. Health. 2006, 209, 31 https://doi.org/10.1016/j.ijheh.2005.08.001
  5. Tsai, J.; Kaye, W. E.; Bove, F. J. Int. J. Hyg. Environ. Health. 2006, 209, 57 https://doi.org/10.1016/j.ijheh.2005.09.003
  6. Sohoni, P.; Lefevre, P. A.; Ashby, J.; Sumpter, J. P. J. Applied Toxico. 2001, 21, 173 https://doi.org/10.1002/jat.747
  7. Roy, S.; Roy, S.; Sharma, C. B. Biomed. Chromato. 2004, 48, 648
  8. Sanchez, R.; Vazquez, A.; Altamirano, J. V.; Villen, J. J. Sci. Food. Agric. 2006, 86, 129 https://doi.org/10.1002/jsfa.2267
  9. Sabater, C.; Carrasco, J. M. Environ. Toxicol. 2001, 16, 314 https://doi.org/10.1002/tox.1038
  10. Jimenez, J. J.; Bernal, J. L.; Nozal, M. J. D.; Arias, E.; Bernal, J. Rapid Commun. Mass Spectrome. 2004, 18, 657 https://doi.org/10.1002/rcm.1383
  11. Diaz, A. N.; Pareja, A. G.; Sanchez, F. G. Pestic. Sci. 1997, 49, 56 https://doi.org/10.1002/(SICI)1096-9063(199701)49:1<56::AID-PS487>3.0.CO;2-3
  12. Deger, A. B.; Gremm, T. J.; Frimmel, F. H. Acta Hydrochim. Hydrobiol. 2000, 28, 292 https://doi.org/10.1002/1521-401X(200012)28:6<292::AID-AHEH292>3.0.CO;2-I
  13. Codognoto, L.; Tanimoto, S. T.; Pedrosa, V. A.; Suffredini, H. B.; Machado, S. A. S.; Avaca, L. A. Electroanal. 2005, 00, 1
  14. Acha, V.; Meurens, M.; Naveau, H.; Agathos, S. N. Biotech. Bioengin. 2000, 68, 473 https://doi.org/10.1002/(SICI)1097-0290(20000605)68:5<473::AID-BIT1>3.0.CO;2-8
  15. Manisankar, P.; Selvanathan, G.; Viswanathan, S.; Prabu, H. G. Electroanal. 2002, 14, 1722 https://doi.org/10.1002/elan.200290016
  16. Diaz, T. G.; Cabanillas, A. G.; Salinas, F. Electroanal. 1999, 12, 616
  17. Garrido, E. M.; Lima, J. L. F. C.; Matos, C. D.; Borges, M. F. M.; Brett, A. M. O. Electroanal. 2001, 13, 199 https://doi.org/10.1002/1521-4109(200103)13:3<199::AID-ELAN199>3.0.CO;2-G
  18. Shi, H,; Zhao, Z.; Song, Z.; Huang, J.;Yang, Y.; Anzai, J.; Chen, T. O. Q. Electroanal. 2005, 17, 1285 https://doi.org/10.1002/elan.200403223
  19. Lin, M. S.; Wang, J. S. Electroanal. 2004, 16, 904 https://doi.org/10.1002/elan.200302896
  20. Snejdarkova, M.; Svododova, L.; Nikolelis, D. P.; Wang, J.; Hianik, T. Electroanal. 2003, 15, 1185 https://doi.org/10.1002/elan.200390145
  21. Koblizek, M.; Masojidek, J.; Komenda, J.; Kucera, T.; Pilloton, R.; Matto, A. K.; Giardi, M. T. Biotech. Bioeng. 1998, 60, 664 https://doi.org/10.1002/(SICI)1097-0290(19981220)60:6<664::AID-BIT3>3.0.CO;2-B
  22. Espinosa, M.; Atanasov, P.; Wikins, E. Electroanal. 1999, 1055
  23. Bond, A. M.; Mahon, P. J.; Schiewe, J.; Victoria, V. B. Anal. Chim. Acta. 1997, 645, 67
  24. Masawat, P.; Liawruangrath, S.; Vaneesorn, Y.; Liawruangrath, B. Talanta. 2002, 58, 1221 https://doi.org/10.1016/S0039-9140(02)00424-1
  25. Perdicakis, M. Aubriet, H.; Walcarius, A. Electroanal. 2004, 16, 2042 https://doi.org/10.1002/elan.200403056
  26. Vestergard, M.; Kerman, K.; Tamiya, E. Analytica Chimica Acta, 2005, 538, 273 https://doi.org/10.1016/j.aca.2005.01.067
  27. Gao, W.; Song, J.; Wu, N. J. Electroanal. Chem. 2005, 5746, 1
  28. Mousty, C. Applied Clay Science, 2004, 27, 159 https://doi.org/10.1016/j.clay.2004.06.005
  29. Wang, J.; Hocevar, S. B.; Ogorevc, B. Electrochem. Commun. 2004, 6, 176 https://doi.org/10.1016/j.elecom.2003.11.010
  30. Korbut, O.; Buckova, M.; Tarapcik, P.; Labuda, J.; Grundler, P. Electroanal. Chem. 2001, 506, 143 https://doi.org/10.1016/S0022-0728(01)00494-6
  31. Radi, A. Talanta, 2005, 68, 271
  32. Ly, S. Y. Bioelectrochem. 2005, 68, 232