According to the Para. 5 of Art 2 of the Korean Nuclear Safety Regulations, which was revised in 1999, internal dose assessment as well as external one should be performed by law for employees at a nuclear power plant from 2003, and their estimate errors should also be within 50%. Thus, more accurate internal dosimetry becomes important. Corresponding to such regulation revision, we are developing a more accurate thyroid-uptake internal dosimetric system and have developed a Monte Carlo simulation code, the so-called CALEFF, to calculate the detection efficiency of the dosimetric system. In this paper, we calculated detection efficiencies with various test conditions by using the CALEFF code and discussed their characteristics. We may use the detection efficiency calculated by the code in calibrating the thyroid internal dose from measured data.