Abstract

We realized the national standard of the candela, one of the SI units, by using two photometers with the spectral responsivity measured in reference to the absolute cryogenic radiometer. The external apertures of the photometers were fabricated using a diamond turning machine, and measured in terms of area with uncertainty of 0.05 %(k = 1). The candela is realized using a 1 kW FEL lamp and the characterized photometers on an optical bench. The uncertainty is budgeted to be 0.25 %(k = 1) considering the uncertainty of the spectral responsivity and the response uniformity of the detectors, the area of the external apertures, the color temperature of the lamp, and the positioning reproducibility of the photometers and the lamp. We verified the realized scale by comparing with the scale of National Institute of Standards and Technology, USA. They coincided with each other within 0.1%.

국제단위계 중의 하나인 칸델라 눈금을 두 개의 광도계를 사용하여 실현하였다. 광도계의 분광감응도는 광출력을 전기출력으로 치환하여 측정하는 극저온 절대복사계에 소급하여 측정하였다. 광도계 외부에 설치된 정밀 개구는 다이아몬드 회전가공기로 가공하였으며 그 면적은 불확도 0.05%(k = 1) 이내로 측정하였다. 이 광도계를 사용하여 광도 벤치 위에서 1 kW FEL 전구로 칸델라 단위를 실현할 때 불확도를 분광감응도, 감응도의 공간 균일도, 개구 면적, FEL 전구의 색온도와 위치 재현성을 고려하여 평가한 결과 0.25%(k = 1)를 얻었다. 실현한 눈금의 유효성을 확인하기 위하여 미국 NIST의 눈금과 비교한 결과 0.1% 이내에서 일치하였다.

Keywords

References

  1. Principles of Light Measurement, Commission Internationale de I'Eclairage(CIE) Pub. no. 18(E-12), 1970
  2. E. Ikonen, P. Karha, ALassila, F. Manoonchehri, H. Fagerlund and L. Liedquist, 'Radiometric realization of the candela with a trap detector,' Metrologia, vol. 32, pp. 689-692, 1996 https://doi.org/10.1088/0026-1394/32/6/59
  3. Albert Parr, 'The candela and photometric and radiometric measurements,' Research Journal of the National Institute of Standards and Technology, vol. 106, pp. 151-186, 2000
  4. Yoshi Ohno, Handbook of Applied Photometry, (edited by Casimer DeCusatis, AIP, New York, 1997) pp. 55-98
  5. J. E. Martin, N. P. Fox, and P. J. Key, 'A cryogenic radiometer for absolute radiometric measurements,' Metrologia, vol. 21, pp. 147-155, 1985 https://doi.org/10.1088/0026-1394/21/3/007
  6. A. Lassila, P. Toivanen, and E. Ikonen, 'An optical method for direct determination of the radiometric aperture area at high accuracy,' Measure. Sci. Technology, vol. 8, pp. 973-977, 1997 https://doi.org/10.1088/0957-0233/8/9/003
  7. 강창호, 김석원, 박승남, '레이저 스폿의 칼날주사 방법에 의한 복사계 개구 면적의 정밀측정,' 한국광학회지, 제15권 제3호, pp. 258-262, 2004 https://doi.org/10.3807/KJOP.2004.15.3.258
  8. BIPM, 'Calibration and Measurement Capabilities of National Metrology Institute,' (http://kcdb.bipm.org/AppendixC)