HRGC/HRMS에 SP-2331 및 DB-5MS 컬럼 사용시 다이옥신 결과의 비교

Comparison of Dioxin Analytical Results Between SP-2331 and DB-5MS Capillary Columns in the HRGC/HRMS Analysis

  • 홍종기 (한국기초과학지원연구원, 유해물질분석연구팀) ;
  • 양정수 (연세대학교, 화학과) ;
  • 신정화 (한국기초과학지원연구원, 유해물질분석연구팀) ;
  • 안윤경 (한국기초과학지원연구원, 유해물질분석연구팀) ;
  • 이대운 (연세대학교, 화학과)
  • Hong, Jongki (Hazardous Substance Research Team, Korea Basic Science Institute) ;
  • Yang, Jeong-Soo (Department of Chemistry, Yonsei University) ;
  • Shin, Jung-Hwa (Hazardous Substance Research Team, Korea Basic Science Institute) ;
  • Ahn, Yun-Gyong (Hazardous Substance Research Team, Korea Basic Science Institute) ;
  • Lee, Dai-Woon (Department of Chemistry, Yonsei University)
  • 투고 : 2003.09.15
  • 심사 : 2003.10.16
  • 발행 : 2004.04.25

초록

본 연구에서는 HRGC/HRMS에 의한 다이옥신분석에 있어 가장 널리 사용되고 있는 SP-2331 및 DB-5MS 컬럼에 의한 독성 2,3,7,8-다이옥신 동족체 분리의 영향에 대하여 조사하였다. 컬럼분리성능평가 표준용액과 비산재 시료의 분석을 통하여 이들 컬럼에 의한 tetra, penta 및 hexa-2,3,7,8-동족체의 분리에 대한 부분과 I-TEQ값의 영향을 중심으로 설명하였다. 이들 두 컬럼에 의한 분석에서 컬럼분리성능평가 표준용액의 경우 DB-5MS에 의해서는 총 2,3,7,8-치환이성체의 농도는 508.4 ng/mL로 나타났으며, SP-2331에 의해서는 515.8 ng/mL로 나타났다. 이들을 I-TEQ값으로 환산하였을 때 거의 동등하게 나타났다. 비산재 시료의 경우에서도 이들 두 컬럼에 의한 2,3,7,8-치환이성체의 총 농도와 I-TEQ값이 거의 비슷하게 나타났다.

In this study, the separation of toxic dioxin 2,3,7,8-congeners by DB-5MS and SP-2331 GC columns which are widely used in HRGC/HRMS analysis was examined. Through the dioxin analysis of column performance check standard solution and fly ash sample, the isomer specific separation of 2,3,7,8-substituted dioxins from tetra to hexa-isomers on DB-5MS and SP-2331 columns were studied. The effect of I-TEQ value by these columns was also studied. The total concentrations of toxic dioxins for the column performance check standard solution were 508.4 ng/mL analyzed by DB-5MS and 515.8 ng/mL analyzed by SP-2331, respectively. The I-TEQ value obtained by both columns was shown to be almost equivalent for the column performance check standard solution and fly ash sample.

키워드

참고문헌

  1. EPA Method 1613, 'Tetra Through Octa-CDDs/Fs by Isotope Dilution HRGC/HRMS', Sept. 1994, Washington, DC.
  2. J. J. Ryan, H.B.S. Conacher, L.G. Panopio, B.P.Y. Lau, J.A. Hardy, Y. Masuda, J. Chromatogr. A, 541, 131-183(1991).
  3. E. Abad, J. Caixach, J. Rivera, J. Chromatogr. A, 786, 125-134(1997).
  4. J. de Boer, J. Chromatogr. A, 843, 179-198(1999).
  5. H.R. Buser and C. Rappe, Anal. Chem., 52, 2257(1980).
  6. 김종국, 박진수, 김경심, 이근희, 김신조, 한국분석과학회지, 15(1), 43-53(2002)
  7. T. Mazer, F.D. Hileman, R.W. Noble, J.J. Brooks, Anal. Chem., 55, 104(1983).