The consideration about exact set-up with stereotactic radiosurgery for lung cancer.

폐암 환자의 전신 정위적 방사선 수술시 정확한 SET UP에 대한 고찰

  • Seo, Dong-Rin (Department of Radiation Oncology, Asan Medical Canter) ;
  • Hong, Dong-Gi (Department of Radiation Oncology, Asan Medical Canter) ;
  • Kwon, Kyung-Tea (Department of Radiation Oncology, Asan Medical Canter) ;
  • Park, Kwang-Ho (Department of Radiation Oncology, Asan Medical Canter) ;
  • Kim, Jung-Man (Department of Radiation Oncology, Asan Medical Canter)
  • 서동린 (서울 아산병원 방사선 종양학과) ;
  • 홍동기 (서울 아산병원 방사선 종양학과) ;
  • 권경태 (서울 아산병원 방사선 종양학과) ;
  • 박광호 (서울 아산병원 방사선 종양학과) ;
  • 김정만 (서울 아산병원 방사선 종양학과)
  • Published : 2004.09.01

Abstract

Purpose : What confirm a patient's set-up precisely is an important factor in stereotactic radiosurgery Especially, the tumor is moved by respiration in case of lung cancer. So it is difficult to confirm a exact location by L-gram or EPID. I will verify a exact patient's set-up about this sort of problem by verification system(exactrac 3.0) Materials and Methods : The patient that had lung cancer operated on stereotactic radiosurgery is composed of 6 people. The 5 patients use an ABC tool and 1 patient doesn't use it. I got such a patient's L-gram and EPID image by Body frame(elekta, sweden), compared Ant. image with Lat. one, and then confirmed a set-up. I fused DRR image of CT and X-ray image of Verification system(exactrac 3.0) 3 dimensional, analyzed the coordinate value(vertical, longitudinal, lateral), and then confirmed a difference of existing method. Results : In case of L-gram and EPID, we judge an exact of the patient's set-up subjectively, and on we could treat the patient with radiation. As a result of using Verification system(exactrac 3.0), coordinate value(vertical, longitudinal, lateral) of patient's set-up was comprised within 5mm. We could estimate a difference of the coordinate value visually and objectively. Consequently, Verification system(exactrac 3.0) was useful in judging an exact patient's set-up. Conclusion : In case of Verification system(exactrac 3.0), we can confirm an exact patient's set-up at any time whenever, However, there are several kinds of the demerit. First, it is a complex process of confirmation than the existing process. Second, thickness of CT scan slice is within 3mm. The last, X-ray image has to have shown itself clearly. If we solve this problem. stereotactic radiosurgery will be useful in treating patients why we can confirm an exact patient's positioning easily.

목적 : 전신 정위적 방사선 수술환자 치료 시 환자의 set up을 정확히 확인하고 평가하는 것은 치료 성적을 좌우 하는데 중요한 요소이다. 특히 폐암 환자의 경우 호흡 등에 의한 종양의 움직임으로 인해서 L-gram이나 EPID를 이용하여 정확한 위치를 확인하는데는 여러 가지 어려움이 있다. 이러한 문제점들에 대하여 verification system(exactrac)을 이용하여 set up의 정확성을 확인해 보고자 한다. 대상 및 방법 : Body frame(elekta, sweden)을 이용하여 전신 정위적 방사선 수술을 받는 폐암 환자 6명을 대상으로 (ABC tool을 사용한 환자 5명과 사용 안한 환자1명) 기존의 L-gram과 EPIDs로 image를 획득하여 Ant, Lat image를 비교하여 환자의 set-up을 확인하고 CT에서 얻은 DRR image와 Verification system(exactrac)의 X-ray image를 얻어서 두image를 3차원적으로 fusion하여 그 좌표 값(vertical, longitudinal, lateral)을 분석하고 기존의 방법과의 차이를 확인하였다. 결과 : L-gram과 EPIDs로는 정확한 좌표 값보다는 주관적으로 Set up의 정확성 여부를 판단하여 치료할 수 있었지만 Verification system(exactrac)을 이용하여 확인한 결과 치료환자의 Set up 전, 후의 좌표 값(vertical, longitudinal, lateral)이 0.5cm이내에 포함되는 것을 볼 수 있었고 좌표 값의 차이를 시각적이고 객관적으로 평가할 수 있어서 Set up의 정확성을 판단하는 데 유용했다. 결론 : Verification system(exactrac)을 사용하면 환자의 Set up이 정확한 지 치료 전 또는 중간에도 시각적인 확인이 가능한 장점이 있지만 기존의 방법보단 확인과정이 조금 복잡하고 CT scan 시 3mm 이내로 해야 하며 X-ray image가 선명하게 나타나야 된다는 단점이 있다. 이러한 문제점만 해결된다면 Set up의 정확성을 쉽게 확인할 수 있어서 치료 시 많은 도움이 되리라고 생각된다.