Transferring Distance-Amplitude Correction Curves - A Model-based Approach

  • Kim, Hak-Joon (Center for NDE, Iowa State University) ;
  • Schmerr Lester W. (Center for NDE, Iowa State University) ;
  • Song, Sung-Jin (School of Mechanical Eng., Sungkyunkwan University) ;
  • Sedov Alexander (Dept. of Mechanical Eng., Lakehead University)
  • 발행 : 2003.12.30

초록

In practice, it is common to manufacture reference blocks containing simple reflectors to obtain distance-amplitude correction (DAC) curves. However, the construction or DAC curves in this manner requires the use of a large number of specimens with appropriate curvatures and reference reflectors located at various depths. Therefore, less costly and quantitative procedures are strongly needed. To address such a need, in this study, we have developed model-based transfer curves to relate a DAC curve obtained in a particular reference configuration with that for a completely different configuration. An example of transferring DAC curves, using the proposed transfer curves, is given.

키워드

참고문헌

  1. Thompson, R. B. and Gray, T. A. (1983) A model relating ultrasonic scattering measurements through liquid-solid interfaces to unbounded medium scattering amplitudes, Journal of Acoustical Society of America, Vol. 74, pp. 1279-1290
  2. Sedov, A., Schmerr, L. W., and Song S. J (1992) Ultrasonic Scattering by a Flat-bottom Hole in Immersion Testing: An Analytical model, Journal of Acoustical Society of America, Vol. 92, pp. 478-486
  3. Schmerr, L. W. (1998) Fundamentals of Ultrasonic Nondestructive Evaluation - A Modeling Approach, Plenum Press, New York
  4. Schmerr, L. W. and Thompson, D. O. (1997) The Role of Modeling in NDE standards, European-American Workshop, Determination of Reliability and Validation of NDE, Germany, pp. 321-328
  5. Auld, B. A (1979) General Electromechanical Reciprocity Relations Applied to the Calculations of Elastic Wave Scattering Coefficients, Wave Motion, Vol. 1, pp. 3-10
  6. Schmerr, L. W and Sedov, A. (2003) Modeling ultrasonic problems for the 2002 Benchmark session, in Review of Progress in Quantitative Nondestructive Evaluation, Vol. 22, eds. D. O. Thompson and D. E. Chimenti, AlP, New York, pp. 1776-1783
  7. Schmerr, L. W. (2000) A Multi-Gaussian Ultrasonic Beam Model for High Performance Simulations on a Personal Computer, Materials Evaluations, Vol (58), pp.882-888
  8. Schmerr, L. W., Song J. S. and Zhang. H. (1994) Model-Based Calibration of Ultrasonic System Responses for Quantitative Measurements, Nondestructive Characterization of Materials VI, pp. 111-118
  9. Rogers, P. H and Van Buren, A. I. (1974) An Exact Expression for the Lommel Diffraction Correction Integral, Journal of Acoustical Society of America, Vol. 55, pp. 724-728
  10. Thompson, R. B. (2002) An Ultrasonic Benchmark Problem: Overview and Discussion of Results, in Review of Progress in Quantitative Nondestructive Evaluation, Vol. 21, eds. D. O. Thompson and D. E. Chimenti, AlP, New York, pp. 1917-1924
  11. Schmerr, L. W. (2002) Ultrasonic Modeling of Benchmark Problems, in Review of Progress in Quantitative Nondestructive Evaluation, Vol. 21, eds. D. O. Thompson and D. E. Chimenti, AlP, New York, pp. 1933-1940