DOI QR코드

DOI QR Code

Calibration and Uncertainty Measurement of Differential Mobility Analyzer Using 100 nm NIST SRM 1963

100 nm NIST 표준입자를 이용한 미분형 전기 이동도 분석기의 교정 및 불확실도 측정

  • 이상진 (한양대학교 대학원 기계공학과) ;
  • 안진홍 (한양대학교 대학원 기계공학과) ;
  • 안강호 (한양대학교 기계공학과)
  • Published : 2003.12.01

Abstract

House made differential mobility analyzer(DMA) is calibrated with NIST SRM 1691(300 nm PSL). Then the particle size and uncertainty for differential mobility analyzer(DMA) using the NIST SRM 1963(100 nm PSL). In result, calibration of prototype DMA is measured using 300 nm NIST SRM 1691, then sheath air flow was corrected 126.67 ㎤/s. Corrected sheath air flow is used in uncertainty measurement of prototype DMA. Uncertainty analysis is performed using NIST SRM 1963(100 nm PSL). The experimental result shows that NIST SRM 1963 is measured as 102.17 nm with a type A uncertainty of 0.33 nm.

Keywords

References

  1. Hewitt, G.W., 1957,Trans. Am. Inst. Elect. Engrs76,p.300
  2. Bademosi, F., 1971, 'Diffusion Charging and Related Transfer Process in Knudson Aerosol,' Ph. D. Thesis, University of Minnesota, Minnesota
  3. Liu, B. Y. H. and Pui, D. Y. H., 1974, 'A submicron Aerosol Standard and the Primary, Absolute Calibration of the Condensation Nuclei Counter,' J. Colloid and Interface Sci., Vol. 47, No. 1, pp. 155-171 https://doi.org/10.1016/0021-9797(74)90090-3
  4. George W. Mulholland, Nelson P. Bryner, Carroll Croarkin, 1999, 'Measurement of the 100nm NIST SRM 1963 by Differential Mobility Analysis,' Aerosol Science and Technology, Vol. 31, pp. 39-55 https://doi.org/10.1080/027868299304345
  5. Kinney, P.D., Pui, D.Y.H., Mulholland, G.W., and Bryner, N.P., 1991,'Use of the Electrical Classification Method to Size $0.1{\mu}m$ SRM Particles-A Feasibility Study,' J. Res. Natl.Inst.Stand.Technol., Vol. 96, pp.147-176 https://doi.org/10.6028/jres.096.006
  6. Oihana Elzalde, Gracia P. Leal, Jose R. Leize, 2000, 'Particle Size Distribution Measurement of Polymeric Dispersions : A Comparative Study,' Part.Part.Syst.Charact., Vol. 17, pp.236-243 https://doi.org/10.1002/1521-4117(200012)17:5/6<236::AID-PPSC236>3.0.CO;2-0
  7. Masa-aki Kitajima, Y. Fukai, 1999, 'Measurement of Size and Number Concentration of Polystyrene Particles,' Colloid Surface A: Physiochemical and Engineering Aspects, Vol. 153, pp. 467-469 https://doi.org/10.1016/S0927-7757(98)00535-4
  8. Kangho Ahn, Namhyo Kim, Jongho Lee, Gwinam Bae, 1996, 'Particle Path and Performance Evaluation of Differential Mobility Analyzer,' Trans. Of the KSME, Vol. 20, No. 6, pp. 2005-2013
  9. G. P. Reischl, 1991, 'Measurement of Ambient Aerosols by the Differential Mobility Analyzer Method: Concepts and Realization Criteria for the Size Range Between 2 and 500nm,' Aerosol Science and Technology, Vol. 14, pp. 5-24 https://doi.org/10.1080/02786829108959467
  10. Barry N. Taylor and Chris E. Kuyatt, 1994, 'Guideline for Evaluating and Expressing the Uncertainty of NIST Measurement Results,' NIST Technical Note 1297
  11. Yongan Gu, Dongqing Li, 1998, 'Measurements of the Electric Charge and Surface Potential on Small Aqueous Drops in the Air by Applying the Millikan Method,' Colloid Surface A: Physiochemical and Engineering Aspects, Vol. 137, pp. 205-215 https://doi.org/10.1016/S0927-7757(97)00366-X
  12. Trans. Of the KSME v.20 no.6 Particle Path and Performance Evaluation of Differential Mobility Analyzer Kangho Ahn;Namhyo Kim;Jongho Lee;Gwinam Bae
  13. Aerosol Science and Technology v.14 Measurement of Ambient Aerosols by the Differential Mobiliy Analyzer Method: Concepts and Realization Criteria for the Size Range Between 2 and 500nm G.P.Reischl https://doi.org/10.1080/02786829108959467
  14. NIST Technical Note 1297 Gyideline for Evaluating and Expressing the Uncertainty of NIST Measurement Results Barry N. Taylor;Chtis E. Kuyatt
  15. Colloid Surface A: Physiochemical and Engineering Aspects v.137 Measurement of the Electric Charge and Surface Potential on Small Aqueous Drops in the Air by Applying the Millikan Method Yongan Gu;Dongqing Li https://doi.org/10.1016/S0927-7757(97)00366-X