Acceptance Test and Quality Control of Radiotherapy Treatment Simulator

의료용 모의치료기의 품질관리에 관한 연구

  • 신동호 (경기대학교 물리학과) ;
  • 박성용 (국립암센터 양성자치료센터) ;
  • 신동오 (경희대학병원 치료방사선과) ;
  • 최진호 (가천의대 길병원 방사선종양학과) ;
  • 김유현 (고려대학교 보건대학 방사선과) ;
  • 권수일 (경기대학교 물리학과)
  • Published : 2002.06.01

Abstract

The simulator which has a identical geometry with radiotherapy equipments, is a essential equipment to accomplish optimized radiotherapy plan through simulation by using diagnostic low energy X-ray. A Radiotherapy simulator has a combined technology from the therapeutic and diagnostic radiology and needs a periodical test for mechanical and optical properties, X-ray generator, image intensifier of simulator to keep the proper maintenance and radiation safety. Hence, tests were done and classified as i) mechanical and optical parameter for the gantry, collimator, and couch ii) key performance of the X-ray generator such as a kVp, mAs, and timer iii) performance of the image intensifier such as a resolution and contrast for three kinds of simulator, common use in clinic. The above result of tests will be applied to the acceptance test and periodical quality assurance procedure.

모의치료기는 방사선치료장비와 기하학적인 구조가 동일하고, 진단용 저 에너지 X-선을 사용한 모의치료과정을 통해 최적화된 방사선치료계획을 수립하는데 필수적인 장비이다. 즉, 진단 및 치료방사선 분야의 기술이 통합되어 있어 모의치료기의 품질관리 시 의료분야의 방사선 안전관리 및 모의치료기의 최적 상태를 유지 관리하기 위해서는 방사선발생장치 요소와 모의치료기의 기하학 및 기계적인 검사 항목들에 대한 주기적인 검사가 필요하다. 따라서 현재 임상에서 사용되고 있는 세 종류의 모의치료기에 대한 갠트리, 콜리메이터 및 환자테이블에 대한 기하학적 및 기계적인 주요 검사 항목과 관전압, 관전류, 타이머 특성 등에 대한 X-선 발생장치의 주요 핵심 항목 및 분해능과 대조도 등의 투시영상증배관의 성능검사로 구분하여 수행하였다. 이를 토대로 모의치료기의 인수검사 및 주기적인 품질관리에 응용하고자 한다.

Keywords