Development of Birdcage RF coil for 3T Animal MR Imaging
Abstract
We authors developed a new small-size birdcage RF coil for animal MR images. And we compared signal-to-noise ratio (SNR) of the new small coil with a conventional knee coil. The dimension of the low-pass type birdcage coil with 12 elements at 37 MRI system are 13 cm outer diameter, 12 cm inner diameter and 20 cm length. For each element, the width of copper tape is 0.05 mm, thickness is 8 mm and length is 20 cm. The small birdcage coil with 12 elements exhibited 7 resonance modes. The isolation of the quadrature channel could be achieved more than 20 ㏈. The coil quality factor (Q value) was 98.6. The SNR of the animal coil was 243.2 on the average and was about twice as high as the conventional knee coil. The present study successfully demonstrated that the small birdcage coil could provide high quality animal MR images with the improved SNR. Therefore, it is expected that the small birdcage coil could be used in the clinical diagnosis and research studies for veterinary medicine in the near future.
목적 : 사람에 비하여 크기가 현저히 작은 동물의 자기공명영상 촬영을 위하여 상용으로 인체에 적용하는 Head 코일보다 원통 반지름이 작은 Low-Pass Type Birdcage 코일을 제작함으로써 보다 큰 신호대잡음비 (Signal-to-Noise Ratio. SNR)를 얻고자 하였다. 방법 : 자기공명영상을 얻기 위해 Spin Echo 펄스시퀀스와 Fast Spin Echo 펄스시퀀스를 사용하였다. T1 강조영상을 얻기 위한 Spin Echo 펄스시퀀스의 매개변수는 TR/TE=300/17 ㎳, Matrix=256