The Journal of Korean Society for Radiation Therapy (대한방사선치료학회지)
- Volume 14 Issue 1
- /
- Pages.59-64
- /
- 2002
- /
- 1598-8449(pISSN)
Effect of skin dose by materials located in treatment field
방사선 치료 시 조사야 내에 위치할 수 있는 이물질이 체표선량에 미치는 영향
- Hong, Chae-Seon (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University, School of Medicine, Seoul, Korea) ;
- Kim, Kyung-Tae (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University, School of Medicine, Seoul, Korea) ;
- Ju, Sang-Gyu (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University, School of Medicine, Seoul, Korea) ;
- Kim, Jong-Sik (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University, School of Medicine, Seoul, Korea) ;
- Park, Young-Hwan (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University, School of Medicine, Seoul, Korea)
- 홍채선 (성균관의대 삼성서울병원 치료방사선과) ;
- 김경태 (성균관의대 삼성서울병원 치료방사선과) ;
- 주상규 (성균관의대 삼성서울병원 치료방사선과) ;
- 김종식 (성균관의대 삼성서울병원 치료방사선과) ;
- 박영환 (성균관의대 삼성서울병원 치료방사선과)
- Published : 2002.12.01
Abstract
Purpose : In radiotherapy, various materials are used to located in treatment field unintentionally. It increases the dose delivered to the skin by interactions of the X-ray within the materials and occurs unwanted skin reaction.(due to the dose build-up effect) This aim of the this study is to measure the increase in skin dose when 13 materials are located in treatment field. Methods : Photon beam measurements were made using an plane-parallel chamber (Markus, PTW-Freiburg) in a polystyrene phantom. skin dose were measured using various overlaying 13 materials. a fixed geometry of a