A Study on Overexposure Rate according to Overdensity in Chest X-ray Radiography(II)

흉부촬영에서 overdensity에 따른 overexposure rate를 아는 방법(II)

  • Kim, Jung-Min (Junior College of Health Sciences, Korea University) ;
  • Huo, Joon (Junior College of Health Sciences, Korea University) ;
  • Hayashi, Taro (The Chemotherapeutic Institute Hospital, Japan)
  • Published : 2000.06.30

Abstract

We have presented with the "A study on overexposure rate according to over-density in chest X-ray radiography(I)" last year. In this report, We could calculate the entrance skin dose from chest X-ray film density the formula $I_0=Ix/e^{-{\mu}x}{\times}mG$, (mG is Bucky factor) was used to deliver the skin dose. At that time, There was two problems that the Bucky factor from maker was not equal to field experience and the field size influenced on the Attenuation Rate. The experiment of Bucky factor was done from film method and retried the Attenuation Rate of Acryle phantom according to Good & Poor geometry. As the results, The Bucky factor from maker higher than in this experiments $30{\sim}40%$. The Attenuation Rate in good geometric condition brings about a little alteration compare with poor geometric condition. In the field experiment, we could get the chest image with very low entrance skin radiation dose $29.3{\mu}Sv$, especially with air gap methode, the entrance skin dose was detected $10{\mu}Sv$.