대한치과보철학회지 (The Journal of Korean Academy of Prosthodontics)
- 제38권1호
- /
- Pages.85-97
- /
- 2000
- /
- 0301-2875(pISSN)
- /
- 2005-3789(eISSN)
임플랜트 고정체와 지대나사간의 부식특성에 관한 연구
CORROSION CHARACTERISTICS BETWEEN IMPLANT FIXTURE AND ABUTMENT SCREW
- Kee, Su-Jin (Department of Prosthodontics, College of Dentistry, Chosun University) ;
- Kweon, Hyeog-Sin (Department of Prosthodontics, College of Dentistry, Chosun University) ;
- Choe, Han-Cheol (Department of Iron-manufacture and metallogical engineering, Kwang-Yang University)
- 발행 : 2000.02.01
초록
The purpose of this study was to compare the corrosion characteristics between implant fixture and two types of abutment screw ; gold screw, titanium screw. The anodic polarization behavior, the galvanic corrosion behavior, and the crevice corrosion behavior of prepared samples were investigated using potentiostat and scanning electron microscope. The results were as follows: 1. Anodic polarization behavior of samples; The primary passivation potential of implant fixture was -420mV, implant abutment was -560mV. titanium screw was -370mV and gold screw was -230mV. All samples were shown to have a high corrosion potential and good formation of passive film. The critical passive current density of gold screw was higher than that of other samples and the sample of gold screw showed a unstable passive film formation at passive region. 2. Galvanic corrosion behavior of samples; Contact current density between implant fixture and titanium screw showed
키워드