Abstract
An absorption spectrometry coupled with flow injection analysis has been developed and tested for the analysis of the Cr$_2O_7^{2-}$/Nn$O_4^-$ mixtures. Even though one has to inject the sample twice into the FIA system and the process of the sample treatment is required to completely destroy the Mn$O_4^-$ ion for the analysis of the Cr$_2O_7^{2-}$ ion, the new method has definite advantages over the current method. It utilizes only a single analytical wavelength (570 nm) and enables one to construct calibration curves which accurately follow the Beer's law over wide ranges of analytical concentrations of both ions ($2.0{\times}10^{-6}$M∼$8.0{\times}10^{-3}$M for Cr$_2O_7^{2-}$ ion, $2.0{\times}10^{-6}$M∼$4.0{\times}10^{-3}$M for MnO4- ion).
산성조건에서 HCr$O_4^-$이온과 $H_2O_2$사이의 산화-환원반응의 생성물인 $CrO(O_2)_2$의흡광도의 측정에 근거를 두어 개발된 흡수분광분석법에 FIA기법을 접목하여 $Cr_2O_7^{2-}$이온과 Mn$O_4^-$이온을 함유하는 혼합시료에 대한 새로운 분석법을 개발하였다. 새로운 분석법은 혼합물 시료의 분석과정에서 두 번에 걸친 주입분석을 수행해야 하고 $Cr_2O_7^{2-}$이온의 분석에 앞서 Mn$O_4^-$이온의 제거를 위한 별도의 작업을 선행해야 한다는 번거로움을 내포하지만, 두 가지의 분석파장들을 사용해야 하는 기존의 흡수분광분석법에 비해 단일분석파장(570 nm)를 사용할 수 있다는 장점을 가진다. 또한, 새로운 분석법은 주어진 분석파장(570 nm)에서 넓은 농도범위($Cr_2O_7^{2-}$이온:$2.0{\times}10^{-6}$M∼$8.0 {\times}10^{-3}$M,Mn$O_4^-$이온:$2.0{\times}10^{-6}$M∼$4.0{\times}10^{-3}$M)에 걸쳐 Beer의 법칙을 정확히 따르는 검정선의 작성이 가능하다는 점에 있어서 기존의 흡수분광분석법에 비해 $Cr_2O_7^{2-}$이온과 Mn$O_4^-$이온의 분석의 용이성을 더욱 증대시킬 수 있다는 장점도 가진다.