A study on the development of oxygen measurement device for diagnosis of peripheral vascular disease in lower extremity

하지 (下脂) 조직내의 말초 혈관계 질환 진단을 위한 산소 측정장치의 개발에 관한 연구

  • 임현수 (충남대학교 의과대학 의공학교실) ;
  • 이준규 (충남대학교 의과대학 정형외과학교실) ;
  • 박동철 (충남대학교 공과대학 전파공학과)
  • Published : 1999.03.01

Abstract

The oxygen saturation of blood can be measured by the difference absorption in optical spectra of Hb and Hb0$_2$, as the well known previous study. In this study we developed the non-invasive oxygen measurement device for diagnosis of peripheral vascular disease in lower extremity using infrared and red LED which produce a peak spectral emission at a wavelength of 660 nm, and 940 nm. To evaluate the clinical application of the oxygen measurement device, we performed lower extremity study to measure the oxygen changes in response to physiological changes within biological tissue. The results showed that oxygen saturation of blood in biological tissue can be monitored from the separation arrangement light source and detector.

혈액에서의 산소 포화도는 Hb와 HbO$_2$의 광학 스펙트라에서 흡수계수의 차이에 의해서 측정할 수 있다. 본 연구는 하지 조직에서 발생하는 말초혈관계 질환진단을 위하여 파장이 660nm 와 940nm의 LED를 이용하여 산소포화도를 측정하는 장치를 개발하였다. 산소측정장치는 광 탐촉자와 광 신호처리부, LED 구동회로, 컴퓨터와 인터페이스로 구성하고, 데이터의 수집과 분석을 위한 프로그램 을 개발하였다. 구현된 산소측정장치에 대한 임상적인 평가를 위하여 하지조직에서 운동 부하에 따른 실험을 하여 생체조직내 생리적변화에 따른 산소량의 변화를 측정하고, 기기의 성능을 평가하였다. 실험결과 생체조직내의 산소포화도는 광원과 검출기의 간격에 따라 측정이 가능함을 보여 주었다.