A Survey on Enterence Dose by Exposure Factors

X-선촬영 조건에 따른 피폭선량 조사연구

  • Kim, Sung-Soo (Dept. of Radiologic Technology, Shin Gu College) ;
  • Huh, Joon (Institute of Radiological Technology, Dong A X-ray co.,LTD.)
  • 김성수 (신구대학 방사선과) ;
  • 허준 (동아엑스선기계 방사선기술연구소)
  • Published : 1998.12.30

Abstract

By analyzing each part of expsure condition and the status of an entrance dose which is exposed to patients, this study reached the following conclusion. Since there is no standard in setting up an exposure condition, the technology practiced in each facility varies tremendously, and the entrance dose increased especially due to the improper selection of screen and grid and the shortage of a total amount of filtration in leaching the standard filtration amount. Entrance dose was, generally, turned out to be $2{\sim}3$ times as much as that of advanced countries, and there was big difference between facilities ; approximately 52 times inlateral of cervical vertebrae, 35 times in A-P of femur, 33 times in chest A-P, and 11 times in lumber A-P. Therefore, to minimize the entrance dose of a patient with thegreatest amount of image information, acquirement of technological know-how necessary for standardization of exposure condition for each part can be an important research task.