Spectrofluorimetric Determination of n-Octanol Based on its Ternary Complex with $Eu^{3+}$ and TTA

$Eu^{3+}$, TTA, 그리고 n-Octanol의 삼성분착물에 의한 n-Octanol의 분광형광분석법에 관한 연구

  • Received : 1997.07.15
  • Published : 1997.12.25

Abstract

The enhancing effect of n-octanol on the fluorescence intensity of the $Eu^{3+}$-thenoyltrifluoroacetone(TTA) system in the presence of Triton X-100 was studied using spectrofluorometric method. This complex exhibited very intense $Eu^{3+}$ ion fluorescence at 619nm, when optically excited at 345nm. Optimum conditions for the determination of n-octanol have also been investigated. The calibration graph was linear over the range $1{\times}10^{-5}M{\sim}1{\times}10^{-7}M$ and the detection limit for n-octanol is $1{\times}10^{-9}M$. The result obtained in the analysis of the synthetic sample agreed with the known value in the error range and the relative standard deviation was ca. 3.5%.

Triton X-100 계면할성제 존재하에서, $Eu^{3+}$-thenoyltrifluoroacetone(TTA)계의 형광세기의 n-octanol 영향에 관해 연구하였고, n-octanol의 정량을 위한 최적 조건을 규명하였다. 이 착물계에서 들뜨기 파장을 345nm로 주사했을 때, $Eu^{3+}$ 이온의 최대 형광파장은 619nm에서 나타났다. n-octanol의 검정곡선은 $1{\times}10^{-5}M{\sim}1{\times}10^{-7}M$의 범위에서 직선적인 관계를 얻었고, 검출한계는 $1{\times}10^{-9}M$이었다. 이 방법을 이용하여 합성시료를 분석한 결과 분석오차내에서 기지값과 일치하였으며, 상대표준편차는 약 3.5%였다.