Investigative Magnetic Resonance Imaging
- Volume 1 Issue 1
- /
- Pages.75-78
- /
- 1997
- /
- 2384-1095(pISSN)
- /
- 2384-1109(eISSN)
Dual Birdcage RF Coil for Leg MR Angiography
하지 MR Angiography를 위한 Dual Birdcage RF 코일
Abstract
A dual birdcage RF coil is proposed for MRI and MRA of the human leg. The proposed coil c can be used to cover the whole human leg by imaging upper and lower parts of the leg separately. In the conventional leg imaging scheme using a single RF coil, the leg has to be r relocated when changing the position of the RF coil thus causing problems in matching the i images from two sets of coils. When using the proposed dual RF coil, however, only the bed h has to be moved to select the imaging region while only one part of the dual coil is used each t time by using current switching and PIN diodes. The utility of proposed coil has been verified b by volunteer MRI and MRA and the imaging results show that the coil is useful for the clinical MRI and MRA of the leg.
인체의 하지 MR Angiography(MRA)를 위한 하지전용 dual birdcage RF 코일을 설계, 제작하였으며 Volunteer실험을 하여 그 유용성을 보였다. 제안된 코일은 인체의 하지 전체의 영상을 위한 코일로 허벅지로부터 종아리까지 전체를 두 번의 영상을 통해 cover하게 된다. 기존의 방식대로 한 개의 코일을 옮기며 두 번 촬영할 경우 코일을 도중에 옮기게 외어 다리를 움직일 수밖에 없어 두 set의 영상이 완전히 연결이 되지 않지만 제안된 코일의 경우 침대만 움직이게 되므로 이런 문제가 없게 된다. 제안된 코일의 성능은 Volunteer의 하지 MRI 및 MRA를 얻어 확인했으며 임상적으로 매우 유용한 것으로 확인되었다.