The Performance Test of Teledyne PB-3 Personnel Dosimetry System by Intercomparison Study

국제상호비교검증을 통한 Teledyne PB-3 개인선량계시스템의 성능시험

  • Published : 1994.08.30

Abstract

Recently, the Ministry of Science and Technology issued a Ministerial Ordinance (No 1992-15) about the technical criteria on personnel radiation dosimetry. In today's climate, it is important to demonstrate and document that the processor's systems and services to others meet national standards of quality. The purpose of this study is to verify the performance of the Teledyne PB-3 personnel dosimetry system that is generally used in Korea Atomic Energy Research Institute(KAERI) by intercomparison with Oak Ridge National Laboratory. The KAERI has been participated in this personnel dosimetry intercomparison study(PDIS) program since 1991 and it could be possible to test and calibrate personnel monitoring system. This report presents a summary and analysis of by about 50 dose equivalent measurements reported for PDIS-16 through 18 (1991 -1993) with emphasis on neutron dose equivalent sensitivity, accuracy and precision. Relationships of the PDIS results to occupational neutron monitoring and methods to improve personnel dosimetry performance are also discussed.

최근, 개인방사선피폭선량평가에 관한 기술기준이 과학기술처에 의해 고시됨에 따라 국내에서도 개인선량계를 이용한 체외피폭선량평가결과의 신뢰성 문제가 크게 대두되고 있다. 한국원자력연구소에서는 이러한 국내의 상황을 인식하고 자체적인 신뢰성 확보를 위해 미국의 Oak Ridge국립연구소의 주관하에 개인피폭선량의 평가에 관한 국제상호비교검증시험(Personnel Dosimet Intercomparison Study ; PDIS)을 수행하였다. 비교검증시험에는 한국원자력연구소에서 사용하고 있는 Teledyne PB-3 열형광선량계가 사용되었으며 선량계산 알고리즘은 Teledyne PB-3 version 1.5-1989를 사용하였다. 본 연구에서는 지금까지 실시 된 PDIS의 결과를 요약하고 현재 PB-3 시스템의 개인선량평가성능에 대해 고찰하였으며, 선량평가절차에 대한 문제점 도출을 통하여 직업적 방사선피폭선량 평가능력의 향상을 위한 방안을 제시하였다.