The Analysis of the Information Capacity of Digital Radiography System

디지탈 X-선 촬영 시스템의 정보용량 분석

  • 김종효 (서울대학교 의과대학 의공학 연구소) ;
  • 민병구 (서울대학교 의과대학 의공학교실, 서울대학교 대학원 전자공학과)
  • Published : 1991.03.01

Abstract

In this paper, the information capacity, the intergrate performance measure of digital radiography system has been quantitatively analyzed. The effect of various factors affecting the information capacity of the digital radiography system in multistage detection processes has been considered and analyzed In detail. The results show that the more information capacity can be obtained with the smaller detector cell area, despite of the reduced signal to noise ratio due to the reduced number of X ray photons per detector cell. Considering the limit of human visual acuity, however. the sufficient resolution will be obtained when the detector cell size in $0.2{\times}0.2\textrm{mm}^2$ with 8 bit quantizaion. And also the results indicates that the information capacity may be severely reduced by the mixture of electric noise in final read-out stage.

Keywords

References

  1. 전자공학회 논문지 v.26 no.7 디지탈 X-선 촬영 시스템의 영상 및 물리적 특성분석 김종효;이태수;박광석;한만청;이충응;민병구
  2. 의공학회지 v.7 no.1 디지탈 X-선 촬영 시스템에 관한 연구 민병구;박광석;이태수;김종효;한만청;임정기
  3. Image Science Image assessment by information theory J.C.Dainty;R.Shaw
  4. Physics in Medicine and Biology no.4 The information spectrum as measure of radiographic image quality and system performance Kanamori(et al.)
  5. Medical Imaging Albert Macovski
  6. 과학기술처 제출 최종보고서 디지탈 래디오그래피 시스템의 개발에 관한 연구 민병구;박광석;김종효
  7. Vision in man and machine Martin D. Levine