A Study on ARC Therapy of 18MeV Linear Accelerator

18MeV 선형가속기의 ARC Therapy에 관한 고찰

  • Kim, Sung-Kyu (Department of Therapeutic Radiology, College of Medicine, Yeungnam University) ;
  • Shin, Sei-One (Department of Therapeutic Radiology, College of Medicine, Yeungnam University) ;
  • Kim, Myung-Se (Department of Therapeutic Radiology, College of Medicine, Yeungnam University)
  • 김성규 (영남대학교 의과대학 치료방사선과학교실) ;
  • 신세원 (영남대학교 의과대학 치료방사선과학교실) ;
  • 김명세 (영남대학교 의과대학 치료방사선과학교실)
  • Published : 1988.12.30

Abstract

In recent years there has been a growing interest in all forms of rotational therapy, and many different types of therapy machines designed for this kind of treatment have become available. To the medical radiation physicist, the dosimetry of rotation therapy has presented a number of interesting problems, and much useful work has been published on the basic data of dose distribution and dosage calculation. The setting dose for ARC therapy were obtained by computer calculation and measurement with cylindrical phantom. Authors compared computer calculation with measured value. And in ARC therapy, the region of maximum dose is shifted from the tumor center. The extent of shift was analyzed by isodose distribution for ARC therapy techniques.

회전조사시 회전각도에 따른 계산치와 측정치를 비교, 검토하여 마음과 같은 결과를 얻었다. $90^{\circ}$ 회전시에는 계산치와 측정치와는 4.5%의 차이를 나타내었고 $120^{\circ}$ 회전시에는 4.4%, $180^{\circ}$ 회전시에는 3.9%, $240^{\circ}$ 회전시에는 2.9%, $360^{\circ}$ 회전시에는 2.1%의 차이를 나타내어 회전각도가 클수록 계산치와 측정치와의 차이는 적은 것으로 나타났다. 또한 회전각도에 따라 최고선량지점이 상부로 이동하는 정도는 $90^{\circ}$에서 $240^{\circ}$까지 회전할 때 tumor depth가 4.5cm일 경우에 는 1.67cm에서 1.10cm 정도 상부로 이동하였으며, tumor depth가 8cm일 경우에는 2.45cm에서 1.40cm 정도 상부로 이동하는 것으로 나타났다.