Abstract
Liquid chromatographic behavior of several metal ions in dithiocarbamate(DTC) chelates were investigated by reversed phase high performance liquid chromatography on Novapak $C_{18}$ and ${\mu}$-Bondapak $C_{18}$ columns. The optimum conditions for the separation of DTC-metal chelate were examined with respect to the pH, shaking time, flow rate, extraction solvent, and mobile phase strength. The metal ions in mixtures at trace level, chelated with some dithiocarbamate derivatives were separated successfully on Novapak $C_{18}$ column using acetonitrile/methanol/water or acetonitril/water mixtures as mobile phases. It was found that all DTC metal chelates studied were eluted in an acceptable range of capacity factor values ($0{\leqq}log\;k'{\leqq}1$). Although several foreign metal ions were coexisted, high recovery and good precision were attained ; 97.0-106.7 % for the recovery and 0.98-3.41% for the coefficient of variation. Under the optimum analytical conditions, trace metal ions in the composite water samples were determined sucessfully with in relative error of about {\pm}$6.7 %.
역상 액체 크로마토그래피에 의하여 몇 가지 금속-dithiocarbamate(DTC) 킬레이트의 용리거동을 Novapak $C_{18}$과 {\mu}$-Bondapak $C_{18}$ 분리관을 사용하여 연구하고, 아울러 동시 분리정량에 미치는 pH, 진탕시간, 흐름속도, 추출용매 종류 및 이동상의 세기등 여러가지 인자들의 영향을 조사 검토하였다. 미량 금속-DTC 킬레이트들은 Novapak $C_{18}$ 분리관에서 acetonitrile/methanol/water 또는 acetonitrile/water의 용리액을 사용하여 성공적으로 분리되었다. 모든 금속-DTC 킬레이트의 용매세기 인자는 $0{\leqq}log\;k'{\leqq}1$ 의 범위임을 확인하였고, 회수율을 97.0-106.7 %, coefficient of variation은 0.98-3.41%이었다. 최적 분석 조건에서 합성 시료중에 있는 혼합 금속 이온들은 상대 오차 ${\pm}$6.7 % 이내에서 동시 분리정량이 가능하였다.