A Study on Dose Distribution around Fletcher-Suit Colpostat Containing $^{137}Cs$ Source

$^{137}Cs$ 선원을 내포한 Fletcher-Suit Colpostat 주위의 선량분포에 관한 연구

  • Kang Wee-Saing (Department of Therapeutic Radiology, College of Medicine, Seoul National University)
  • 강위생 (서울대학교 의과대학 치료방사선과학교실)
  • Published : 1988.12.01

Abstract

This paper presents dose distributions in water around Fletcher-Suit colpostat containing $^{137}Cs$ tube, and shielding effect of Internal lead shield. Using ready packed film, author measured dose distribution in water around the colpostat containing cesium source. Nine sheets of films on one side of the colpostat are packed with acryl frames cut out so as to fill water, and irradiated in water by cesium source in the colpostat. Dose distributions on transverse plane and upper plane 0.5cm from upper surface of the colpostat were measured. Shielding effect was greater in upper medial direction than in lower medial direction. And that was the greatest around $30^{\circ}$ from the axis of the colpostat on upper side and around $50^{\circ}$on lower side. In the region 7cm from the center of the colpostat, shielding efficiency was 0.23 to 0.35 on the lower $50^{\circ}$ and 0.26 to 0.42 on the upper $30^{\circ}$, and decreased with increase of distance.

[ $30^{\circ}$ ] 선원을 내포한 Fletcher-Suit Colpostat주위의 물에서 선량 분포와 내부의 납차폐물에 의한 차폐 효과가 본 논문에 기술된다. 미리 포장된 필름을 이용하여 세슘선원을 내포한 콜포스타트 주위의 선량 분포를 측정하였다. 콜포스타트의 한 방향에 9매의 필름을 아크릴 틀을 이용하여 평행하게 고정시키고 세???讚㈉湲??? 이용하여 조사하였다. 콜포스타트의 내측과 외측은 동시에 조사하였으며, 전방은 따로 조사하였다. 콜포스타트의 종축이 포함된 내외 횡단면과 윗쪽 표면 밖 0.5cm 떨어 진 만면에서 선량분포를 얻었다. 차폐효율은 콜포스타트의 상부측이 하부측보다 켰으며, 상부측에서는 종축에서 내측으로 $30^{\circ}$정도의 선과 하부측에서는 $50^{\circ}$정도의 선에서 차폐효율이 각각 가장 큰 것으로 나타났다. 콜포스타트의 중심에서 반경 7cm범위내에서 차폐효율은 상부 $30^{\circ}$에서는 0.26-0.42, 하부 $50^{\circ}$에서는 0.23-0.35였으며 거리가 밀어짐에 따라 감소하였다. 콜포스타트 상부표면에서 0.5cm떨어진 만면에서 대략 3cm 반경의 원내에서는 차폐효과가 있었지만 그 밖에서는 거의 없었다.

Keywords