Charateristics of 10MV X-ray Beam from a Mevatron KD Linear Accelerator

Mevatron KD 선형 가속기에서의 10MV X-선 특성

  • Yi, Byong-Yong (Department of Radiation Therapy, Hangyang University Hospital) ;
  • Lee, Myung-Za (Department of Radiation Therapy, Hangyang University Hospital)
  • 이병용 (한양대학교 의과대학 치료방사선과학교실) ;
  • 이명자 (한양대학교 의과대학 치료방사선과학교실)
  • Published : 1988.06.01

Abstract

The beam characteristics and dosimetric measurements of the 10MV X-ray beam from a Mevatron KD linear accelerator are examined. The Percent Depth Dose (POD) table and the Tissue Maximum Ratio (TMR) table are taken from measurement as a function of the field size and the depth. The calculated TMR table from PDD table is compared with those from measurement. Other beam characteristics such as output factor, beam profile (including flatness, symmetry and penumbra), wedge, and the variation of Dmax are presented.