Abstract

X-ray high-voltage generator is the most important part that can decide the radiation exposure dose affecting a patient or operator according to the characteristic. If decrease of X-ray radiation exposure dose and output characteristic of high-voltage generator is unstable, a patient or operator must be exposed to more radiation. This study measures and analyzes the exposure dose reproducibility and output characteristic according to a change of tube current on the various rectification methods of diagnostic X-ray equipment. It can find that quality bastardize and output is increased if voltage of X-ray tube is increased. Exposure dose reproducibility according to output of X-ray equipment is extremely excellent in inverter type, and is stable in order of following three-phase, a single-phase and condenser method. This study can find that the reply incidence of high-voltage generator is generated due to difference in rectification method, noise occurs in X-ray due to that, quality of an image is decreased due to that, and medical diagnosis can be failed due to that.

Keywords