대한전기학회:학술대회논문집 (Proceedings of the KIEE Conference)
- 대한전기학회 2008년도 제39회 하계학술대회
- /
- Pages.1977-1978
- /
- 2008
생체 전위 측정에서 2-전극 차동 증폭 시스템과 2-전극 비차동 증폭 시스템의 비교
Comparison between a differential and a non-differential amplifier system with two electrodes in bio-potential measurement
- Kang, Dae-Hun (Yonsei University) ;
- Lee, Chung-Keun (Yonsei University) ;
- Lee, Sang-Joon (Yonsei University) ;
- Lee, Myoung-Ho (Yonsei University)
- 발행 : 2008.07.16
초록
In this paper, we compare performance of common-mode rejection between a differential and a non-differential amplifier system with two electrodes. A differential amplifier system is constant for common-mode rejection ratio(CMRR) on the frequency domain. But a non-differential amplifier's CMRR is determined by
키워드