Measurement of MTF in Display image on monitor by SWRF

SWRF를 이용한 전시영상의 MTF 측정

  • Kim, Chang-Bok (Dept. of Electrical and Electronic Eng., Dong Shin University) ;
  • Lee, Kyung-Sup (Dept. of Electrical and Electronic Eng., Dong Shin University) ;
  • Kim, Young-Keun (Dept. of Radiological Kwang-ju Health Collage) ;
  • Cho, Su-Young (Dept. of Electrical Control and Instrumentation Eng., Sung Hwa College)
  • 김창복 (동신대학교 전기전자공학과) ;
  • 이경섭 (동신대학교 전기전자공학과) ;
  • 김영근 (광주보건대학 방사선과) ;
  • 조수영 (성화대학 전기제어계측과)
  • Published : 2004.11.19

Abstract

The quality of X -ray image is depending on the acquired signal pattern So far, the MTF(modulation transfer function) of medical image system has been obtained through various kinds of calculationprocesses after using SWRF (A square wave response function method) to print out it into films. In this study, a tool has been developed that can help the actual user who actually creates the medical images measure the MTF of the finaldisplay image very simply by applying Borland C++ builder software as well as LEAD tools software for the SWRF calculation process to analyze the MTF even on the display image. Films have been printed out by this newly developed MTF measuring tool under the same conditions with the ones for the existing tools and also the calculation method has shown no difference with any existing SWRF calculation method. By using this, it was found out that the MTF of the resolving power of the computed radiography(CR) Image plate (IP) that is a conventional X-ray detection system Jar the conventional X-ray purpose was about 10% at the range of 3.00 LP/mm.