Computer Simulation of Double Osteotomy in Tibia Using Double-Hexapod External Fixator

이중 헥사포드 외고정 기구를 이용한 선천성 왜소증 환자의 경골 교정 시뮬레이션

  • Published : 2004.10.01

Abstract

In this paper, we proposed a new method of pre-operative planning for tibial deformity correction using double hexapod external fixator in pseudoacondroplasia. The 3-D computer graphic model of deformed tibia was reconstructed from 3 mm sliced CT data, and CAD model of double hexapod external fixator was developed. The fixator was composed of 170 mm diameter of three rings and 90 mm of twelve struts. The bone deformities and the osteotomy lines for double osteotomy were measured using X-rays, and the necessary joint values to correct the given deformities were obtained by inverse kinematics analysis. The computer graphic simulation was performed to visualize the deformity correction process and evaluate the analysis result. By examining the pre-op and post-op X-rays, the simulation result was in good agreement with the clinical outcomes.