Application of an Energy Sensitive CZT Detector to a DXA Type of Bone Densitometer

  • Yoon, Je-Woong (ISOL Technology Inc., Dept. of Biomedical Engineering, The catholic University of Korea) ;
  • Lee, Hyung-Koo (Dept. of Biomedical Engineering, The catholic University of Korea) ;
  • Lee, Heung-Kyu (ISOL Technology Inc., Dept. of Biomedical Engineering, The catholic University of Korea)
  • 발행 : 2002.09.01

초록

The accuracy of DXA(Dual Energy X-ray Absorptiometry) highly depends on the detection and separation capability of dual energy X-ray X-ray photons. In addition both of scan time and patient exposure are affected by detection efficiency. A CZT detector with a good energy resolution and high detection efficiency was evaluated for the application of bone densitometry. Its performance was compared to a photomultiplier tube with a NaI(T1) scintillator in terms of energy resolution, detection efficiency and the accuracy of bone mineral density measurement. The comparison study was performed with CZT detector and PM tube using DXA equipments(OSTEO Plus, OSTEO Prima, ISOL Technology). The energy spectrum was acquired using MCA(Multi-Channel Analyzer). The used X-ray energy ranged from 20keV to 86keV. The MCA result of the CZT detector showed a slightly sharper energy spectrum than that of NaI(T1). Detection efficiency of the CZT detector at 59.5keV was 1.4 times better. Remarkably the final results of bone mineral density measurements demonstrate only less than 1% difference. The CZT detector appears to have many benefits for the application of bone densitometry. Its excellent energy resolution can enhance the counting accuracy of dual energy X-ray spectrum. Furthermore its compactness in physical dimension and no cooling requirement will be additional benefits for a more compact and accurate bone densitometer.

키워드